EasyFerm Plus PHI MS 225
Found in these product categories:
Part/REF # 238633-1443
-
(In Stock)(In Stock)Stock status is updated daily. For real-time status, please contact us.
Specifications
Sensor Family |
Sensor Family
EasyFerm Plus
|
---|---|
Parameter |
Parameter
pH
|
Sensor Output |
Sensor Output
Memosens
|
a-length |
a-length
225 mm
|
Electrical Connector |
Electrical Connector
Memosens
|
Measurement Principle |
Measurement Principle
Combination electrode; pH potential measured against reference
|
Measuring Range |
Measuring Range
pH 0 to 14
|
Additional specifications
Sensitivity |
Sensitivity
57 to 59 mV / pH at 25 °C
|
---|---|
Zero Point |
Zero Point
0 ± 20 mV
|
pH Glass |
pH Glass
Hamilton Type PHI Glass
|
pH Glass Resistance |
pH Glass Resistance
Moderate
|
Membrane Shape |
Membrane Shape
Spherical
|
Diaphragm |
Diaphragm
HP-Coatramic
|
Number of Diaphragms |
Number of Diaphragms
1
|
Electrolyte |
Electrolyte
Pressurized Phermlyte
|
Reference System |
Reference System
Hamilton Everef-F, Silver ion barrier included
|
Liquid Earth |
Liquid Earth
No
|
Temperature Sensor |
Temperature Sensor
NTC 30 kOhm
|
Diameter |
Diameter
12 mm
|
Process Connection |
Process Connection
PG13,5
|
Wetted Parts |
Wetted Parts
Glass
Ceramic EPDM (Ethylene propylene elastomer) |
O-ring Material |
O-ring Material
FDA 21 CFR 177.2600; EG 1935/2004; USP <87>; USP <88> Class VI (121 °C)
|
Serial Number |
Serial Number
Yes
|
Recommended Storage |
Recommended Storage
Hamilton Storage Solution or 3 M KCl
|
Certificate |
Certificate
Yes, with measured values in Buffer pH 4.01 and pH 7.00
|
ATEX Approval |
ATEX Approval
II 1 G Ex ia IIC T3/T4/T6 Ga
|
IECEx Approval |
IECEx Approval
II 1 G Ex ia IIC T3/T4/T6 Ga
|
Autoclavable |
Autoclavable
Yes
|
CIP |
CIP
Yes
|
Steam Sterilizable |
Steam Sterilizable
Yes
|
Operating Temperature Range |
Operating Temperature Range
0 to 135 °C; 0 to 130 °C (ATEX)
|
Sample min Conductivity |
Sample min Conductivity
100 µS/cm
|
Pressure Range bar g |
Pressure Range bar g
0 to 6 bar
|
Resources
Documents
-
Brochure
-
Catalogs
-
Certificates
- ATEX Certificate BVS 11 ATEX E 038 X - Memosens Sensors (PDF)
- ATEX Certificate TÜV 03 ATEX 7005 X - Analog Sensors and Armatures (PDF)
- BSE / TSE / Jatropha Statement (PDF)
- Certificate - China RoHS Hazardous Substances in Parts (PDF)
- Certificates for Raw Material according the EN 10204-3.1.B
- Declaration of Quality for Calibration Solutions
- EU Declaration of Conformity - Analog Sensors and Armatures (ATEX) (PDF)
- EU Declaration of Conformity - Arc Sensors (PDF)
- EU Declaration of Conformity - Memosens Sensors and Memosens Cables (PDF)
- EU Declaration of Conformity - Memosens Sensors and Memosens Cables (PDF)
- IECEx Certificate - Analog Sensors and Housings (PDF)
- IECEx Certificate BVS 16.0079X - Memosens Sensors (PDF)
- UKCA Declaration of Conformity - Analog Sensors and Armatures (UKEX) (PDF)
-
Material Specifications
-
SDS
-
Technical Drawings
-
Technical Manuals

We're Ready to Help
Hamilton's sales, customer service, and technical support staff are available for questions and additional information about all products and services.
Contact Us