From Ranch to Relish
Benzoic acid and sorbic acid are two of the most routinely monitored preservatives in food and beverage production, but getting a reliable method that works across very different matrices is easier said than done. This application note puts the Hamilton PRP-Z2 polymeric stationary phase to the test, evaluating a single isocratic HPLC method for separating and quantifying both preservatives in nine real-world matrices: wine, ranch dressing, pickles, pickle relish, prepared mustard, hot sauce, jelly, ginger ale, and energy drinks.
The results hold up well across the board. Resolution stayed above 2.0 in every matrix, tailing factors fell between 1.17 and 1.23, and spike recovery came in at 103% plus or minus 1.1%, all within a run time of under six minutes. A separate longevity study using wine pushed the column past 1,000 injections while maintaining consistent calibration and system suitability, making a strong case for the PRP-Z2 in labs where a wide range of sample types and long column life both matter.
Author: Adam L. Moore, PhD, Hamilton Company