Reversed-Phase Columns

Packing NameMaterialExchange CapacityPore SizeTemp LimitsMobile Phase LimitsBuffer StrengthMaximum Pressure

PRP-1

PS-DVB*N/A100 Å5-85 °CpH 1-13, 0-100% aqueous, organic modifier
0.0-0.5 N5,000 psi
PRP-C18
C18 bonded to PS-DVB*
N/A100 Å5-85 °CpH 1-13, 0-100% aqueous, organic modifier
0.0-0.5 N
5,000 psi
PRP-3
PS-DVB*
N/A
300 Å
5-85 °C
pH 1-13, 0-100% aqueous, organic modifier
0.0-0.5 N
5,000 psi
PRP-h5
Fluorinated PS-DVB*
N/A
300 Å
5-85 °C
pH 1-13, 0-100% aqueous, organic modifier
0.0-0.5 N
5,000 psi
HxSil C8SilicaN/A110 Å5-60 °CpH 1.5-10, 0-100% aqueous, organic modifier
N/A5,000 psi
HxSil C18
Silica
N/A
110 Å
5-60 °C
pH 1.5-10, 0-100% aqueous, organic modifier
N/A
5,000 psi

Anion Exchange Columns

Packing NameSupport MaterialExchange CapacityPore SizeTemp. LimitsMobile Phase LimitsBuffer StrengthMaximum Pressure
PRP-X100PS-DVB* with tri-methyl ammonium exchanger0.19 meq/gm100 ÅpH Dependent**pH 1-13. 0-100% aqueous, organic modifier0.0-0.5 N5,000 psi
PRP-X110PS-DVB* with Tri-methyl ammonium Exchanger0.11 meq/gm100 ÅpH Dependent**pH 1-13. 0-100% aqueous, organic modifier0.0-0.5 N5,000 psi
PRP-X500Poly(meth-acryl amido-propyl Trimethyl-ammonium chloride)1.6 meq/gmSuperficially porouspH Dependent**pH 1-13. 0-100% aqueous, organic modifier0.0-2.0 N5,000 psi
PRP-X600(di-methyl amido-propyl meth-acrylamide)1.6 meq/gmSuperficially porouspH Dependent**pH 1-13. 0-100% aqueous, organic modifier0.0-1.0 N5,000 psi
RCX-10PS-DVB* with Tri-methyl ammonium Exchanger0.35 meq/gm100 ÅpH Dependent**pH 1-13. 0-100% aqueous, organic modifier0.0-1.0 N5,000 psi
RCX-30PS-DVB* with Tri-methyl ammonium Exchanger1.0 meq/gm100 ÅpH Dependent**pH 1-13. 0-100% aqueous, organic modifier0.0-1.0 N5,000 psi

Cation Exchange Columns

Packing NameSupport MaterialExchange CapacityPore SizeTemp. LimitsMobile Phase LimitsBuffer StrengthMaximum Pressure
PRP-X200PS-DVB* Sulfonate Exchanger35 ueq/gm100 Å5-60 °CpH 1-13. 0-100% aqueous, organic modifier0.0-0.5 N5,000 psi
PRP-X400PS-DVB* Sulfonate Exchanger2.5 meq/gmN/A5-60 °CpH 1-13. 0-100% aqueous, organic modifier0.0-5.0 N5,000 psi
PRP-X800PS-DVB* Itaconate Exchanger1.6 meq/gm100 Å5-60 °C0-100% aqueous, organic modifier0.0-0.5N5,000 psi
HC-40PS-DVB* Sulfonate Exchanger5 meq/gmGel-type25-90 °C100% WaterWater400 psi
HC-75 H+PS-DVB* Sulfonate Exchanger5 meq/gmGel-type25-90 °C100% Water 0-40% Acetonitrile WaterWater400 psi
HC-75 Ca+2PS-DVB* Sulfonate Exchanger5 meq/gmGel-type25-90 °C100% Water 0-40% Acetonitrile WaterWater400 psi
HC-75 Pb+2PS-DVB* Sulfonate Exchanger5 meq/gmGel-type25-90 °C100% Water 0-40% Acetonitrile WaterWater400 psi

Ion Exclusion

Packing NameSupport MaterialExchange CapacityPore SizeTemp. LimitsMobile Phase LimitsBuffer StrengthMaximum Pressure
PRP-X300PS-DVB* Sulfonate Exchanger0.17 meq/gm100 Å5-60 °CpH 1-13. 0-100% aqueous, organic modifier0.0-0.5 N5,000 psi

*PS-DVB is Poly(styrene-divinylbenzene)

**(5-60 °C, pH 1-7.9) / (5-30 °C, pH 8-13)