Temperature Influence on DuraCal pH Buffer 9.21
pH changes based on the ambient temperature. This article provides the pH values from 5 to 50°C for Hamilton's 9.21 DuraCal pH liquid buffers (Refs 238319, 238219, 238919, 238216, 238897).
A prerequisite for an accurate pH calibration is the availability of suitable buffer solutions. Hamilton 9.21 DuraCal liquid buffer solutions have an accuracy of +/- 0.02 pH over 60 months. Like all aqueous solutions, the pH value of the buffer is affected by changing temperature. The effects of temperature from 5 to 50°C is shown below in 1-degree increments.
pH vs. Temperature: Reference table for 9.21 DuraCal pH buffer
(Refs 238319, 238219, 238919, 238216, 238897)
| Temp [°C] | pH Value | Temp [°C] | pH Value |
|---|---|---|---|
| 5 | 9.45 | 30 | 9.16 |
| 6 | 9.43 | 31 | 9.15 |
| 7 | 9.42 | 32 | 9.14 |
| 8 | 9.41 | 33 | 9.13 |
| 9 | 9.39 | 34 | 9.12 |
| 10 | 9.38 | 35 | 9.11 |
| 11 | 9.37 | 36 | 9.10 |
| 12 | 9.36 | 37 | 9.09 |
| 13 | 9.34 | 38 | 9.08 |
| 14 | 9.33 | 39 | 9.07 |
| 15 | 9.32 | 40 | 9.06 |
| 16 | 9.31 | 41 | 9.06 |
| 17 | 9.30 | 42 | 9.05 |
| 18 | 9.28 | 43 | 9.04 |
| 19 | 9.27 | 44 | 9.03 |
| 20 | 9.26 | 45 | 9.03 |
| 21 | 9.25 | 46 | 9.02 |
| 22 | 9.24 | 47 | 9.01 |
| 23 | 9.23 | 48 | 9.01 |
| 24 | 9.22 | 49 | 9.00 |
| 25 | 9.21 | 50 | 8.99 |
| 26 | 9.20 | ||
| 27 | 9.19 | ||
| 28 | 9.18 | ||
| 29 | 9.17 |