Volume |
|
0.5 µL |
0.5 µL |
1 µL |
1 µL |
2 µL |
2 µL |
5 µL |
Model |
|
7000.50C |
7000.5 |
7001 |
7101 |
7002 |
7102 |
7105 |
Scale length, mm |
|
60 |
60 |
60 |
60 |
60 |
60 |
60 |
Total scale divisions |
|
100 |
100 |
100 |
100 |
100 |
100 |
100 |
Major graduations, µL |
|
0.05 |
0.05 |
0.1 |
0.1 |
0.2 |
0.2 |
0.5 |
Minor graduations, µL |
|
0.01 |
0.01 |
0.02 |
0.02 |
0.04 |
0.04 |
0.1 |
Sub-minor graduations, µL |
|
0.005 |
0.005 |
0.01 |
0.01 |
0.02 |
0.02 |
0.05 |
Maximum temperature |
°C |
115 |
115 |
115 |
115 |
115 |
115 |
115 |
|
°F |
239 |
239 |
239 |
239 |
239 |
239 |
239 |
Minimum temperature |
°C |
10 |
10 |
10 |
10 |
10 |
10 |
10 |
|
°F |
50 |
50 |
50 |
50 |
50 |
50 |
50 |
Maximum pressure rating |
bars |
414 |
414 |
414 |
414 |
414 |
414 |
414 |
|
psig |
6000 |
6000 |
6000 |
6000 |
6000 |
6000 |
6000 |
Standard needle gauge |
|
32 |
25 |
25 |
22 |
25 |
23 |
24 |
Standard needle outer diameter (O.D.) |
mm |
0.235 |
0.515 |
0.515 |
0.718 |
0.515 |
0.642 |
0.566 |
|
inches |
0.0093 |
0.0203 |
0.0203 |
0.0283 |
0.0203 |
0.0253 |
0.0223 |
Standard needle inner diameter (I.D.) |
mm |
0.104 |
0.104 |
0.145 |
0.145 |
0.206 |
0.257 |
0.325 |
|
inches |
0.0041 |
0.0041 |
0.0057 |
0.0057 |
0.0081 |
0.0081 |
0.0128 |
Standard needle length |
mm |
100 |
70 |
70 |
70 |
70 |
70 |
70 |
|
inches |
3.94 |
2.75 |
2.75 |
2.75 |
2.75 |
2.75 |
2.75 |
Standard electro-taper |
|
No |
No |
No |
Yes |
No |
Yes |
No |
Standard needle dead volume, µL |
|
0 |
0 |
0 |
0 |
0 |
0 |
0 |