PRP-X200 10 µm 2.1 x 150 mm
PRP-X200 columns can quickly resolve alkali metals, alkaline earth metals, and ammonium. One unique feature of this column is that the resolution between any two ions in the alkali metal series can be increased or decreased by changing the concentration of methanol in the mobile phase; allowing focus on the particular ion of interest and reducing interference from other ions in the sample.
Application Examples: Inorganic and organic cations, alkali and alkaline earth metals, mono or divalent cations.
Found in these product categories:
Part/REF # 79394-
Stock status is updated daily. For real-time status, please contact us.
Specifications
Separation Mode |
Separation Mode
Cation Exchange
|
---|---|
Stationary Phase |
Stationary Phase
PRP-X200
|
Format |
Format
HPLC Column
|
Hardware Inner Diameter |
Hardware Inner Diameter
2.1 mm
|
Hardware Length |
Hardware Length
150 mm
|
Particle Size |
Particle Size
10 µm
|
USP Classification |
USP Classification
USP L22
|
Additional specifications
Packing Material Type |
Packing Material Type
PSDVB/Sulfonic Acid
|
---|---|
Pore Size |
Pore Size
100 Å
|
Particle Surface Area (m2/cm3) |
Particle Surface Area (m2/cm3)
450
|
Pore Volume (cm3/g) |
Pore Volume (cm3/g)
0.752
|
Hardware Material |
Hardware Material
Stainless Steel
|
Max Pressure |
Max Pressure
5,000 psi
|
Exchange Capacity |
Exchange Capacity
35 µeq/gm
|
Mobile Phase Limits |
Mobile Phase Limits
pH 1-13. 0-100% aqueous, organic modifier
|
Buffer Strength |
Buffer Strength
0.0-0.5N
|
Temp Limits |
Temp Limits
5-60°C
|
Restoration |
Restoration
Inject several times with 100 µL of 1 N nitric acid
|
Compound Classification |
Compound Classification
Organic/Inorganic Cations
|
UOM |
UOM
1 EACH
|
Resources
Documents
-
Guides
-
SDS
Related Content
- HPLC Guard Columns, HPLC Guard Cartridges Article
- How Do I Connect a Stainless Steel Guard Column to My HPLC Column? FAQ
- How Do I Connect a PEEK Guard Cartridge to My HPLC Column? FAQ
- Choosing Column Specifications Article
- HPLC Column Instructions for Polymer Stationary Phases Article
Work with R&D engineers to customize Hamilton parts
Hamilton's "Custom Built" program puts you directly in contact with a Hamilton R&D Engineer to discuss your application and product needs. The engineer will point you to a standard solution or propose a customized solution to meet your needs.
Talk to an engineer