What happens if the cell density in the process is ultimately much higher than the cell density used for the correlation with Dencytee Arc?

The software then interpolates the values but points out this fact. The following options are available to the user:

  • Ignore the hint and trust that the interpolation into the much higher density range is correct.
  • Add another correlation point to the existing correlation with the higher in- and off-line values (if the max. 6 points have not been used already)
  • Modify the 6th point of the correlation and overwrite it with the higher values. This correlation cannot be saved with the same correlation-name, it must be saved under a new name.