What flow rates are possible with the Microlab 600?

The Microlab 600 syringe drive is able to achieve a complete 60 mm stroke in as fast as 1.5 seconds and as slow as 3600 seconds. This translates to different flow rates depending on the syringe and the tubing that is used on the instrument. The table below defines the minimum, maximum, and default flow rate for each syringe volume.

Syringe VolumeRecommended (µL/sec)Minimum (µL/sec)Maximum (µL/sec)
10 µL50.0036.5
25 µL12.50.00716.5
50 µL250.01433
100 µL500.0366.5
250 µL1250.07166.5
500 µL2500.14333
1,000 µL5000.3665
2,500 µL6250.71,250
5,000 µL1,2501.42,500
10,000 µL2,50035,000
25,000 µL3,12576,000*
50,000 µL3,125146,000*

* For these syringes it may be required to cut the tip of the dispense tube off to avoid stalling the instrument.