Passionate about cutting-edge automation, robotics, and laboratory technologies? Explore open positions and accelerate your career with us. Search Job Openings

Is the measurement from Dencytee Arc less reliable if the quality index decreases?

Although the probe is very robust and designed for long process life, it is subject to a certain amount of aging. The quality indicator takes the operating hours and sensor diagnostic results into account and recommends replacement when needed. In addition, the probe has other built-in features that constantly check the sensor, even during a measurement.

The evaluation of the quality monitoring is also included in the Quality Index (QI). This is a precautionary measure to avoid the failure of the sensor in a process. However, the quality of the measurement result is not reduced by a lower QI. Dencytee Arc will still function well and measure reliably below 40%, but the sensor will have a warning that it needs to be replaced soon. When the QI drops to 10%, the sensor will no longer be functioning reliably and will hold an error stating that it needs to be replaced.