Temperature Influence on DuraCal pH Buffer 8.00 (Ref 238277)

Green circle icon with a dark navy lightbulb containing two interlocking gears, on a teal dotted background.

A prerequisite for an accurate pH calibration is the availability of suitable buffer solutions. Hamilton 8.00 DuraCal liquid buffer solution (Ref 238277) has an accuracy of +/- 0.02 pH units and a stability of 60 months. The pH value of the buffer is affected by changing temperature. The effects of temperature from 5 to 50°C is shown below in 1-degree increments.

pH vs. Temperature: Reference table for 8.00 DuraCal pH buffer
 

Temp [°C]pH ValueTemp [°C]pH Value
58.09307.98
68.09317.98
78.08327.98
88.08337.97
98.07347.97
108.07357.97
118.06367.96
128.06377.96
138.05387.96
148.05397.95
158.05407.95
168.04417.95
178.04427.95
188.03437.94
198.03447.94
208.02457.94
218.02467.94
228.01477.94
238.01487.93
248.01497.93
258.00507.93
268.00  
277.99  
287.99  
297.99