A prerequisite for an accurate pH calibration is the availability of suitable buffer solutions. Hamilton 6.00 DuraCal liquid buffer solution (Ref 238276) has an accuracy of +/- 0.02 pH units and a stability of 60 months. The pH value of the buffer is affected by changing temperature. The effects of temperature from 5 to 50°C is shown below in 1-degree increments.

pH vs. Temperature: Reference table for 6.00 DuraCal pH buffer

Temp [°C]pH ValueTemp [°C]pH Value
56.02306.00
66.02316.00
76.02326.00
86.01336.00
96.01346.00
106.01356.00
116.01366.00
126.01376.00
136.01386.01
146.01396.01
156.00406.01
166.00416.01
176.00426.01
186.00436.02
196.00446.02
206.00456.02
216.00466.02
226.00476.03
236.00486.03
246.00496.03
256.00506.04
266.00  