Temperature Influence on DuraCal pH Buffer 3.06 (Ref 238274)

Green circle icon with a dark navy lightbulb containing two interlocking gears, on a teal dotted background.

A prerequisite for an accurate pH calibration is the availability of suitable buffer solutions. Hamilton 3.06 DuraCal liquid buffer solution (Ref 238274) has an accuracy of +/- 0.02 pH units and a stability of 60 months. The pH value of the buffer is affected by changing temperature. The effects of temperature from 5 to 50°C is shown below in 1-degree increments.

pH vs. Temperature: Reference table for 3.06 DuraCal pH buffer

Temp [°C]pH ValueTemp [°C]pH Value
53.12303.05
63.12313.05
73.11323.05
83.11333.05
93.11343.04
103.10353.04
113.10363.04
123.10373.04
133.09383.04
143.09393.04
153.09403.04
163.08413.04
173.08423.04
183.08433.04
193.08443.04
203.07453.04
213.07463.04
223.07473.04
233.07483.04
243.06493.04
253.06503.04
263.06  
273.06  
283.05  