A prerequisite for an accurate pH calibration is the availability of suitable buffer solutions. Hamilton 2.00 DuraCal liquid buffer solution (Ref 238273) has an accuracy of +/- 0.02 pH units and a stability of 60 months. The pH value of the buffer is affected by changing temperature. The effects of temperature from 5 to 50°C is shown below in 1 degree increments.

pH vs. Temperature: Reference table for 2.00 DuraCal pH buffer

Temp [°C]pH ValueTemp [°C]pH Value
51.99301.99
61.99311.99
71.99321.99
82.00331.99
92.00341.98
102.00351.98
112.00361.98
122.00371.98
132.00381.98
142.00391.98
152.00401.98
162.00411.98
172.00421.97
182.00431.97
192.00441.97
202.00451.97
212.00461.97
222.00471.97
232.00481.97
242.00491.97
252.00501.97
262.00  
271.99  
281.99  
291.99