Passionate about cutting-edge automation, robotics, and laboratory technologies? Explore open positions and accelerate your career with us. Search Job Openings

Temperature Influence on DuraCal pH Buffer 2.00 (Ref 238273)

Green circle icon with a dark navy lightbulb containing two interlocking gears, on a teal dotted background.

A prerequisite for an accurate pH calibration is the availability of suitable buffer solutions. Hamilton 2.00 DuraCal liquid buffer solution (Ref 238273) has an accuracy of +/- 0.02 pH units and a stability of 60 months. The pH value of the buffer is affected by changing temperature. The effects of temperature from 5 to 50°C is shown below in 1 degree increments.

pH vs. Temperature: Reference table for 2.00 DuraCal pH buffer

Temp [°C]pH ValueTemp [°C]pH Value
51.99301.99
61.99311.99
71.99321.99
82.00331.99
92.00341.98
102.00351.98
112.00361.98
122.00371.98
132.00381.98
142.00391.98
152.00401.98
162.00411.98
172.00421.97
182.00431.97
192.00441.97
202.00451.97
212.00461.97
222.00471.97
232.00481.97
242.00491.97
252.00501.97
262.00  
271.99  
281.99  
291.99