Temperature Influence on DuraCal pH Buffer 11.00 (Ref 238278)

This article provides the measured pH values from 5 to 50°C for Hamilton's 11.00 DuraCal pH liquid buffer (Ref 238278).

Green circle icon with a dark navy lightbulb containing two interlocking gears, on a teal dotted background.

A prerequisite for an accurate pH calibration is the availability of suitable buffer solutions. Hamilton 11.00 DuraCal liquid buffer solution (Ref 238278) has an accuracy of +/- 0.05 pH units and a stability of 24 months. The pH value of the buffer is affected by changing temperature. The effects of temperature from 5 to 50°C is shown below in 1-degree increments.

pH vs. Temperature: Reference table for 11.00 DuraCal buffer

Temp [°C]pH ValueTemp [°C]pH Value
511.333010.93
611.323110.91
711.303210.90
811.283310.89
911.263410.87
1011.253510.86
1111.233610.85
1211.213710.83
1311.193810.82
1411.173910.81
1511.164010.80
1611.144110.78
1711.124210.77
1811.114310.76
1911.094410.74
2011.074510.73
2111.064610.72
2211.034710.71
2311.034810.70
2411.024910.68
2511.005010.67
2610.99  
2710.97  
2810.96  
2910.94  
3010.93