Passionate about cutting-edge automation, robotics, and laboratory technologies? Explore open positions and accelerate your career with us. Search Job Openings

Temperature Influence on DuraCal pH Buffer 10.01

pH can change based on the temperature.  This article lists the expected pH values from 5 to 50°C for Hamilton's 10.01 DuraCal pH liquid buffers (Refs 238321, 238223, 238923, 238187, 238898).

Green circle icon with a dark navy lightbulb containing two interlocking gears, on a teal dotted background.

A prerequisite for an accurate pH calibration is the availability of suitable buffer solutions. Hamilton 10.01 DuraCal liquid buffer solutions are considered "Certified Reference Material" and are tested by a 3rd party lab - DAkkS - against primary reference solutions from PTB and NIST. 10.01 liquid buffers have an accuracy of +/- 0.02 pH over 60 months. Like all aqueous solutions, the pH value of the buffer is affected by changing temperature. The effects of temperature from 5 to 50°C is shown below in 1-degree increments.

pH vs. Temperature: Reference table for 10.01 DuraCal buffer

(Refs 238321, 238223, 238923, 238187, 238898)

Temp [°C]pH ValueTemp [°C]pH Value
510.19309.97
610.18319.95
710.17329.94
810.17339.93
910.16349.92
1010.15359.92
1110.14369.91
1210.13379.90
1310.13389.89
1410.12399.88
1510.11409.86
1610.10419.86
1710.09429.85
1810.08439.84
1910.07449.83
2010.06459.83
2110.05469.82
2210.04479.81
2310.03489.80
2410.02499.80
2510.01509.79
2610.00  
279.99  
289.98  
299.97