Temperature Influence on DuraCal pH Buffer 10.01

pH can change based on the temperature.  This article lists the expected pH values from 5 to 50°C for Hamilton's 10.01 DuraCal pH liquid buffers (Refs 238321, 238223, 238923, 238187, 238898).

Green circle icon with a dark navy lightbulb containing two interlocking gears, on a teal dotted background.

A prerequisite for an accurate pH calibration is the availability of suitable buffer solutions. Hamilton 10.01 DuraCal liquid buffer solutions are considered "Certified Reference Material" and are tested by a 3rd party lab - DAkkS - against primary reference solutions from PTB and NIST. 10.01 liquid buffers have an accuracy of +/- 0.02 pH over 60 months. Like all aqueous solutions, the pH value of the buffer is affected by changing temperature. The effects of temperature from 5 to 50°C is shown below in 1-degree increments.

pH vs. Temperature: Reference table for 10.01 DuraCal buffer

(Refs 238321, 238223, 238923, 238187, 238898)

Temp [°C]pH ValueTemp [°C]pH Value
510.19309.97
610.18319.95
710.17329.94
810.17339.93
910.16349.92
1010.15359.92
1110.14369.91
1210.13379.90
1310.13389.89
1410.12399.88
1510.11409.86
1610.10419.86
1710.09429.85
1810.08439.84
1910.07449.83
2010.06459.83
2110.05469.82
2210.04479.81
2310.03489.80
2410.02499.80
2510.01509.79
2610.00  
279.99  
289.98  
299.97