Temperature Influence of Conductivity Standard 84 µS/cm (Ref 238984)

Hamilton offers the 84 µS/cm liquid conductivity standard (Ref 238984) with stability of ±1% is guaranteed over a lifecycle of up to 18 months. These standards can be used several times, on condition that the bottle is not left open without its lid for more than one hour total. The composition of these standards is patented, and the procedure for determining conductivity has been developed in collaboration with the Danish Institute of Fundamental Metrology (FDM) and the National Institute of Standards and Technology (NIST). Please refer to the calibration certificate packaged with every Hamilton Conductivity Standard for further details.

Line graph showing conductivity (µS/cm) vs. temperature (°C) for a 84 µS/cm standard, rising from ~53 at 5°C to ~130 at 50°C.


 

Conductivity vs. Temperature: Reference table

Temperature °C

Conductivity µS/cm

Temperature °C

Conductivity µS/cm

Temperature °C

Conductivity µS/cm

552.62177.537104.4
6542279.138106.2
755.62380.739108
857.12482.340110
958.7258441111.7
10602685.642113.5
1161.82787.343115.4
1263.32888.944117.3
1364.92990.645119
1466.4309246121.1
1568319447123
1669.63295.748125
1771.13397.449127
1872.83499.150129
1974.335101  
207636102.7