Temperature Influence of Conductivity Standard 706 µS/cm (Ref 238929)

Conductivity standard 12880us cm 500ml 238988

Hamilton offers several conductivity standards of varied values, whose stability of ±2% is guaranteed over a lifecycle of up to 3 years. These standards can be used several times, on condition that the bottle is not left open without its lid for more than one hour total. The composition of these standards is patented, and the procedure for determining conductivity has been developed in collaboration with the Danish Institute of Fundamental Metrology (FDM) and the National Institute of Standards and Technology (NIST). Please refer to the calibration certificate packaged with every HAMILTON Conductivity Standard for further details.

Line graph showing conductivity (µS/cm) vs. temperature (°C) for a 706 µS/cm standard, rising from ~350 at 5°C to ~1300 at 50°C.

Conductivity vs. Temperature: Reference table

Temperature °C

Conductivity µS/cm

Temperature °C

Conductivity µS/cm

Temperature °C

Conductivity µS/cm

53592162737971
63712264838994
738623666391018
840124686401040
941725706411066
1042926727421091
1144927748431116
1246528769441140
1348229790451166
1449930813461191
1551631834471217
1653432856481242
1755233878491268
1857134901501295
1958935924  
2060936947