Temperature Influence of Conductivity Standard 5
Hamilton is the first manufacturer to offer Conductivity Standards at 5 μS/cm (Ref 238926), with a certified accuracy of ±1% and a 3-year lifetime. To prevent any gas permeation from the atmosphere, which could affect conductivity, this product is specially packaged in a 250mL glass bottle.
The composition of the standards is patented, and the procedure for determining conductivity has been developed in collaboration with the Danish Institute of Fundamental Metrology (DFM) and the National Institute of Standards and Technology (NIST). Please refer to the calibration certificate packaged with every HAMILTON Conductivity Standard for further details.
Conductivity vs. Temperature: Reference table
Temperature °C | Conductivity μS/cm | Temperature °C | Conductivity μS/cm | Temperature °C | Conductivity μS/cm |
|---|---|---|---|---|---|
| 5 | 1.5 | 21 | 4 | 37 | 8.8 |
| 6 | 1.6 | 22 | 4.3 | 38 | 9.1 |
| 7 | 1.7 | 23 | 4.5 | 39 | 9.5 |
| 8 | 1.8 | 24 | 4.7 | 40 | 10 |
| 9 | 2 | 25 | 5 | 41 | 10.4 |
| 10 | 2.1 | 26 | 5.3 | 42 | 10.8 |
| 11 | 2.2 | 27 | 5.5 | 43 | 11.2 |
| 12 | 2.4 | 28 | 5.8 | 44 | 11.7 |
| 13 | 2.5 | 29 | 6.1 | 45 | 12.1 |
| 14 | 2.7 | 30 | 6.4 | 46 | 12.6 |
| 15 | 2.9 | 31 | 6.7 | 47 | 13.1 |
| 16 | 3.1 | 32 | 7 | 48 | 13.6 |
| 17 | 3.2 | 33 | 7.3 | 49 | 14.1 |
| 18 | 3.4 | 34 | 7.7 | 50 | 14.7 |
| 19 | 3.6 | 35 | 8 | ||
| 20 | 3.8 | 36 | 8.4 |
Related Content:
- Return to directory for Temperature Influence on Conductivity Standards
- See our Liquid Calibration Solutions Brochure to learn more about these products
- Learn more about the Stability and Accuracy of Hamilton conductivity calibration liquid standards
- Visit our main Conductivity Knowledge Base Page to see all articles
- See Hamilton Conductivity Standards