Temperature Influence of Conductivity Standard 147 µS/cm (Ref 238985)

ACC COND STAND 147us cm 300ml

Hamilton offers the 147 µS/cm liquid conductivity standard (Ref 238985), whose stability of ±1% is guaranteed over a lifecycle of up to 18 months. These standards can be used several times, on condition that the bottle is not left open without its lid for more than one hour total. The composition of these standards is patented, and the procedure for determining conductivity has been developed in collaboration with the Danish Institute of Fundamental Metrology (FDM) and the National Institute of Standards and Technology (NIST). Please refer to the calibration certificate packaged with every HAMILTON Conductivity Standard for further details.

Line graph showing conductivity (µS/cm) vs. temperature (°C) for a 147 µS/cm standard, ranging 5–50°C.


 

Conductivity vs. Temperature: Reference table

Temperature °C

Conductivity µS/cm

Temperature °C

Conductivity µS/cm

Temperature °C

Conductivity µS/cm

5922113637183
6952213838186
7972314139189
81002414440192
91032514741195
101052615042199
111082715343202
121112815644205
131132915945209
141163016246212
151193116547215
161223216748218
171243317149222
181273417450225
1913035176  
2013336180