Temperature Influence of Conductivity Standard 1413 µS/cm (Ref 238928 Glass Bottle)

Hamilton offers the 1413 µS/cm liquid conductivity standard in a glass bottle, whose stability of ±1% is guaranteed over a lifecycle of up to 3 years. These standards can be used several times, on condition that the glass bottle is not left open without its lid for more than one hour total. The composition of these standards is patented, and the procedure for determining conductivity has been developed in collaboration with the Danish Institute of Fundamental Metrology (FDM) and the National Institute of Standards and Technology (NIST). Please refer to the calibration certificate packaged with every Hamilton Conductivity Standard for further details.

temperature influence chart

Conductivity vs. Temperature: Reference table for 1413 μS/cm (Glass Bottle)

Temperature °C

Conductivity µS/cm

Temperature °C

Conductivity µS/cm

Temperature °C

Conductivity µS/cm

5717211255371944
6746221293381992
7775231333392039
8805241373402088
9836251413412137
10868261455422186
11899271497432235
12932281539442285
13965291582452336
14999301626462387
151033311670472438
161069321714482490
171105331759492542
181142341805502594
191179351851  
201217361897