Temperature Influence of Conductivity Standard 1413 µS/cm (Ref 238986 Calpack Bottle)

Hamilton offers the 1413 µS/cm liquid conductivity standard in a convenient Calpack plastic bottle with stability of ±1% guaranteed over a lifecycle of up to 18 months. These standards can be used several times, on condition that the bottle is not left open without its lid for more than one hour total. The composition of these standards is patented, and the procedure for determining conductivity has been developed in collaboration with the Danish Institute of Fundamental Metrology (FDM) and the National Institute of Standards and Technology (NIST). Please refer to the calibration certificate packaged with every HAMILTON Conductivity Standard for further details.

Line graph showing conductivity (µS/cm) vs. temperature (°C) for 1413 µS/cm Calpack calibration solution, ranging ~900–2150 µS/cm from 5–50°C.

Conductivity vs. Temperature: Reference table for 1413 μS/cm (Plastic Bottle)

Temperature °C

Conductivity µS/cm

Temperature °C

Conductivity µS/cm

Temperature °C

Conductivity µS/cm

5893211305371751
6918221329381780
7943231358391810
8968241385401843
9993251413411869
101018261440421899
111044271468431929
121069281495441959
131095291523451988
141121301552462020
151149311579472051
161173321607482082
171199331636492113
181225341665502145
191251351691  
201278361722