Temperature Influence of Conductivity Standard 12880 µS/cm (Ref 238988)

Hamilton offers the 12880 µS/cm liquid conductivity standard (Ref 238988), whose stability of ±1% is guaranteed over a lifecycle of up to 3 years. These standards can be used several times, on condition that the bottle is not left open without its lid for more than one hour total. The composition of these standards is patented, and the procedure for determining conductivity has been developed in collaboration with the Danish Institute of Fundamental Metrology (FDM) and the National Institute of Standards and Technology (NIST). Please refer to the calibration certificate packaged with every HAMILTON Conductivity Standard for further details.

Line graph showing conductivity (µS/cm) vs. temperature (°C) for a 12880 µS/cm standard, rising from ~8200 at 5°C to ~19400 at 50°C.


 

Conductivity vs. Temperature: Reference table

Temperature °C

Conductivity µS/cm

Temperature °C

Conductivity µS/cm

Temperature °C

Conductivity µS/cm

5821021119103715920
6844022121403816180
7866023124003916440
8888024126404016690
9911025128804116960
10934026131404217220
11956027133804317490
12979028136304417750
131002029138804518020
141025030141404618280
151050031143904718550
161072032146404818820
171096033148904919080
181120034151505019350
19114303515420  
20116703615660