Temperature Influence of Conductivity Standard 100,000 µS/cm (Ref 238935)

Hamilton's highest conductivity liquid standard is a 100,000 μS/cm value with a certified accuracy of ±1% and a lifetime of 3 years. The composition of these standards is patented, and the procedure for determining conductivity has been developed in collaboration with the Danish Institute of Fundamental Metrology (FDM) and the National Institute of Standards and Technology (NIST). Please refer to the calibration certificate packaged with every HAMILTON Conductivity Standard for further details.

Line graph showing conductivity (µS/cm) vs. temperature (°C) for a 100,000 µS/cm standard, ranging 5–50°C.


 

Conductivity vs. Temperature: Reference table

Temperature °CConductivity μS/cmTemperature °CConductivity μS/cmTemperature °CConductivity μS/cm
563,0002192,00037124,000
665,0002294,00038127,000
767,0002396,00039129,000
868,0002498,00040131,000
970,00025100,00041133,000
1072,00026102,00042135,000
1174,00027104,00043137,000
1275,00028106,00044140,000
1377,00029108,00045142,000
1479,00030110,00046144,000
1581,00031112,00047146,000
1683,00032114,00048149,000
1784,00033116,00049151,000
1886,00034118,00050153,000
1988,00035120,000  
2090,00036122,000