Standardize Cell Lysis for Reproducible Downstream Analysis

Manual cell lysis can introduce variability between experiments and interrupt automated cell culture workflows. This application note demonstrates how Cell Care STAR™ integrates cell lysis with automated cultivation and imaging. Tested with HeLa and HEK293 cells using different lysis buffers, the workflow delivered comparable protein recovery to manual processing while reducing plate-to-plate variability by approximately tenfold.

Key Takeaways

  • Reduced variability: Automated lysis reduced plate-to-plate variability by approximately tenfold compared with manual processing.
  • Flexible cell processing: Adjustable lysis parameters and a dedicated Sensitive Cells mode accommodate different cell types and lysis conditions.
  • Integrated sample preparation: Automated lysis connects cell cultivation and imaging with downstream molecular analysis, reducing manual intervention.