VisiTrace mA 120 L1
The next generation VisiTrace mA 120 L1 is designed to measure dissolved oxygen in the low ppb ranges in brewing applications, notably during filtration and filling. In addition, the special designed ODO Cap L1 for breweries is stabilized against standard disinfectant solution with active chlorine and chlorine dioxide. This is powerful during measurements in breweries, which may not allow for calibration after every CIP. With the transmitter integrated, the intelligent VisiTrace mA sensor provides more reliable measurements directly to your process control system via the 4-20 mA output. The also integrated Bluetooth 5 wireless interface may be used for monitoring, configuration, and calibration, and saves time without compromising quality.
Found in these product categories:
Part/REF # 10068709-1111-
(In Stock)(In Stock)Stock status is updated daily. For real-time status, please contact us.
Specifications
Sensor Family |
Sensor Family
VisiTrace mA
|
---|---|
Parameter |
Parameter
DO (Optical)
|
Sensor Output |
Sensor Output
4 to 20 mA / HART
|
a-length |
a-length
120 mm
|
Electrical Connector |
Electrical Connector
M12
|
Measurement Principle |
Measurement Principle
Oxygen dependent luminescence quenching, temperature compensation
|
Measuring Range |
Measuring Range
0 to 2 ppm (0 to 50 mbar)
|
Additional specifications
Accuracy at 25 °C |
Accuracy at 25 °C
± 0.5 ppb or 2 % whichever is greater
|
---|---|
Drift at Room Temperature |
Drift at Room Temperature
< 1 % per week @ 100 ppb
|
Sensor Cap |
Sensor Cap
ODO Cap L1
|
Electrolyte |
Electrolyte
None
|
Oxygen Consumption |
Oxygen Consumption
None
|
Temperature Sensor |
Temperature Sensor
NTC 22 kOhm
|
Configurable Values |
Configurable Values
DO: mbar; %-sat; %-vol; µg/l; mg/l; ppb/ppm (gas); ppb/ppm (dissolved oxygen); Temperature: °C, °F, K
|
Diameter |
Diameter
12 mm
|
Process Connection |
Process Connection
PG13,5
|
Wetted Parts |
Wetted Parts
Stainless Steel 1.4435
EPDM (Ethylene propylene elastomer) Silicone - FDA 21 CFR 177.2600 VMQ (Silicone elastomer) See compliance details in Material Specification document |
Surface Quality of Steel |
Surface Quality of Steel
Ra < 0.4 µm (N5)
|
4 to 20 mA Current Range |
4 to 20 mA Current Range
3.5 to 22 mA
|
Analog Interface 1 |
Analog Interface 1
Two wire sink 4 to 20 mA for DO, programmable
|
4 to 20 mA Accuracy |
4 to 20 mA Accuracy
< 0.3 % + 0.05 mA
|
Transmitting Power |
Transmitting Power
Max. 4 dBm
|
Digital Interface |
Digital Interface
Bluetooth 5.0
HART Version 7.0 |
Operating Voltage |
Operating Voltage
18 to 30 VDC
|
Serial Number |
Serial Number
Yes
|
Response Time |
Response Time
t90% < 20 s in gas; < 90 s in water
|
Certificate |
Certificate
Yes, with measured current in oxygen
|
ATEX Approval |
ATEX Approval
ll 1 G Ex ia llC T3/T4/T5/T6 Ga; II 1 D Ex ia IIIC T135°C Da
|
IECEx Approval |
IECEx Approval
ll 1 G Ex ia llC T3/T4/T5/T6 Ga; II 1 D Ex ia IIIC T135°C Da
|
Autoclavable |
Autoclavable
Yes
|
CIP |
CIP
Yes
|
Steam Sterilizable |
Steam Sterilizable
Yes, max. Temperature 140 °C
|
Operating Temperature Range |
Operating Temperature Range
-10 to 140 °C; the sensor provides no DO reading above 85 °C
|
Pressure Range bar g |
Pressure Range bar g
-1 to 12 bar
|
Protection Rating |
Protection Rating
IP 68
|
Required Flow |
Required Flow
None
|
Resources
Documents
-
Brochure
-
Catalogs
-
Certificates
- ATEX Certificate - Explosion Protection Directive 94/9/EC, Annex IV (PDF)
- ATEX Certificate BVS 13 ATEX E 065 X - VisiPro DO Ex and VisiTrace DO Ex (PDF)
- BSE / TSE / Jatropha Statement (PDF)
- Certificate - China RoHS Hazardous Substances in Parts (PDF)
- Certificates for Raw Material according the EN 10204-3.1.B
- Declaration of Quality for Calibration Solutions
- EU Declaration of Conformity - VisiFerm mA and VisiTrace mA Sensors (ATEX) (PDF)
- EU Declaration of Conformity - VisiPro DO and VisiTrace DO Sensors (PDF)
- IECEx Certificate BVS 13.0075X - VisiPro DO Ex and VisiTrace DO Ex (PDF)
- IEXEx Conformity - VisiFerm mA and VisiTrace mA Sensors (PDF)
- UKCA Declaration of Conformity - VisiFerm RS485-ECS / VisiTrace RS485 Sensors (PDF)
- Wireless Certificate - VisiFerm mA & VisiTrace mA (Canada) (PDF)
- Wireless Certificate - VisiFerm mA & VisiTrace mA (Japan) (PDF)
- Wireless Certificate - VisiFerm mA & VisiTrace mA (USA) (PDF)
- Wireless Certificate - VisiPro & VisiTrace (Canada) (PDF)
- Wireless Certificate - VisiPro & VisiTrace (Japan) (PDF)
- Wireless Certificate - VisiPro & VisiTrace (USA) (PDF)
- Wireless Certificate - VisiTrace mA (China) (PDF)
- Wireless Certificate - VisiTrace mA (India) (PDF)
- Wireless Certificate - VisiTrace mA (South Korea) (PDF)
-
Flyers
-
Material Specifications
-
Software Downloads
-
Technical Drawings
-
Technical Manuals
Related Content
- Is there documentation available to transition from the legacy VisiFerm Arc to the VisiFerm RS485? FAQ
- Does the VisiFerm RS485 have the same form, fit, and function as the legacy VisiFerm Arc and VisiFerm ECS? FAQ
- Will upgrading to the VisiFerm RS485 change my calibration routine? FAQ
- How is sensor configuration and calibration performed using the VisiFerm RS485? FAQ
- How do I connect my VisiFerm RS485 to the ArcAir software? FAQ
We're Ready to Help
Hamilton's sales, customer service, and technical support staff are available for questions and additional information about all products and services.
Contact Us