Needle FAQ Header

Needle Gauge Chart

Use the Hamilton Needle Builder

When selecting needle gauge sizes, it is important to keep in mind the volume of the syringe and the dead volume of the needle. For example, it will be very difficult to prime a 10 µL syringe if the dead volume in the needle is greater than 10 µL. Refer to the needle gauge chart to choose a needle gauge with an appropriate µL/inch before selecting a needle. Select the minimum needle length that allows you to carry out your application comfortably.

Note: The ‘s’ on a 22s needle represents a smaller I.D. (inner diameter) for the needle and a thicker needle wall for better durability. For example, a 26 gauge needle has an O.D. (outer diameter) of 0.46 mm and an I.D. of 0.26 mm, while the 26s gauge needle has an O.D. of 0.47 mm and an I.D. of 0.13 mm. The 26s has half the I.D. of the 26 gauge needle. Also, the difference in the wall thickness nearly doubles, with 26s gauge having a thickness of 0.18 mm while the 26 gauge is only 0.10 mm.

View Hamilton Needles Try the Visual Needle Builder

Common ApplicationsGauge Number  Needle Nominal O.D. (mm)Needle Nominal I.D. (mm)Needle Wall Thickness (mm)Needle Dead Volume (µL/25.4mm)
Life Science34 gauge needle0.1590.0510.0510.052 µL/25.4 mm
33 gauge needle0.210.1080.0510.233 µL/25.4 mm
32 gauge needle0.2350.1080.0640.233 µL/25.4 mm
31 gauge needle0.2610.1330.0640.353 µL/25.4 mm
30 gauge needle0.3120.1590.0760.504 µL/25.4 mm
29 gauge needle0.3370.1840.0760.675 µL/25.4mm
28 gauge needle0.3620.1840.0890.675 µL/25.4 mm
27 gauge needle0.4130.210.1020.876 µL/25.4 mm
Chromatography26s gauge needle0.4740.1270.1780.322 µL/25.4 mm
26 gauge needle0.4640.260.1021.349 µL/25.4 mm
25s gauge needle0.5150.1530.1780.464 µL/25.4 mm
25 gauge needle0.5150.260.1271.349 µL/25.4 mm
24 gauge needle0.5660.3110.1271.93 µL/25.4 mm
23s gauge needle0.6420.1160.2670.268 µL/25.4 mm
23 gauge needle0.6420.3370.1522.266 µL/25.4 mm
22s gauge needle0.7180.1680.2790.563 µL/25.4 mm
22 gauge needle0.7180.4130.1523.403 µL/25.4 mm
General Fluid Handling21 gauge needle0.8190.5140.1525.271 µL/25.4 mm
20 gauge needle0.9080.6030.1527.255 µL/25.4 mm
19 gauge needle1.0670.6860.1919.389 µL/25.4 mm
18 gauge needle1.270.8380.21614.011 µL/25.4 mm
17 gauge needle1.4731.0670.20322.715 µL/25.4 mm
16 gauge needle1.6511.1940.22928.444 µL/25.4 mm
15 gauge needle1.8291.3720.22937.529 µL/25.4 mm
14 gauge needle2.1091.60.25451.076 µL/25.4 mm
13 gauge needle2.4131.8040.30564.895 µL/25.4 mm
12 gauge needle2.7692.1590.30593 µL/25.4 mm
11 gauge needle3.0482.3880.33113.728 µL/25.4 mm
10 gauge needle3.4042.6930.356144.641 µL/25.4 mm
 *mm are nominal    

Looking for Information not found in our Syringe Needle Gauge Chart?

See the Hamilton line of Syringe Needles
Download the PDF version of the Needle Gauge Chart