Incyte Sensor Maintenance
-
Incyte Replacement Sensors
-
Incyte Pre-amp and Calibration Accessories
Part # / Ref # Description 238988 Conductivity standard 12880 µS/cm, Basic Line 243708 Solution B for Incyte Calibrations LC 243713 Validation / Calibration Kit for Incyte LC 243720 Incyte Pre-Amp 243740 Validation / Calibration Kit for Incyte 243742 Solution B for Incyte Calibrations 243743 Permittivity Simulator
Specific Applications
-
App Note - Human Platelet Protein Production from Human Cell Culture using an Advanced Bioreactor
Online viable cell density monitoring in a platelet bioprocess production.
Resources
-
Quick Guide - Incyte Sensors (Pre-amp Version)
Quick Guide for Incyte Sensors (Pre-amp Version) -
Manual - Cell Density Measuring System (Pre-Amp Version with Arc View Controller)
Operating Instructions for the Cell Density Monitoring System for use with Arc View Controllers -
Material Specification - Incyte Cell Density Sensors (Arc and Pre-amp Versions)
Material specification for Incyte sensors and units (both Incyte Arc and Pre-amp products)
-
How often do Incyte sensors (pre-amp version) need to be calibrated?
Hamilton recommends once per year with a validation at 6 months. -
What effect will bubbles have on the Incyte sensor measurement?
Gas bubbles have two main impacts: (1) Decreased signal intensity since the sensor cannot measure in air. (2) Increased background noise. -
What is the effect of agitation variations on the Incyte measurement?
The agitation can influence the size of air bubbles. This number and size of air bubbles can affect the Cell Density measurement.
-
Incyte Sensors (Pre-amp Version) - Standard vs. Low Conductivity
There are several different options for Incyte cell density sensors based on liquid conductivity. Learn more about sensor selection / applications.