Temperature Influence on DuraCal pH Buffer 12.00 (Ref 238279)
A prerequisite for an accurate pH calibration is the availability of suitable buffer solutions. Hamilton 12.00 DuraCal liquid buffer solution (Ref 238279) has an accuracy of +/- 0.05 pH units and a stability of 24 months. The pH value of the buffer is affected by changing temperature. The effects of temperature from 5 to 50°C is shown below in 1 degree increments.
pH vs. Temperature: Reference table for 12.00 DuraCal buffer
Temp [°C] | pH Value | Temp [°C] | pH Value |
---|---|---|---|
5 | 12.46 | 30 | 11.90 |
6 | 12.44 | 31 | 11.88 |
7 | 12.41 | 32 | 11.86 |
8 | 12.39 | 33 | 11.84 |
9 | 12.36 | 34 | 11.82 |
10 | 12.34 | 35 | 11.80 |
11 | 12.32 | 36 | 11.78 |
12 | 12.29 | 37 | 11.76 |
13 | 12.27 | 38 | 11.74 |
14 | 12.25 | 39 | 11.72 |
15 | 12.23 | 40 | 11.70 |
16 | 12.20 | 41 | 11.68 |
17 | 12.18 | 42 | 11.66 |
18 | 12.17 | 43 | 11.64 |
19 | 12.13 | 44 | 11.62 |
20 | 12.11 | 45 | 11.60 |
21 | 12.09 | 46 | 11.58 |
22 | 12.05 | 47 | 11.57 |
23 | 12.04 | 48 | 11.55 |
24 | 12.02 | 49 | 11.53 |
25 | 12.00 | 50 | 11.51 |
26 | 11.98 | ||
27 | 11.96 | ||
28 | 11.94 | ||
29 | 11.92 |