Temperature Influence on DuraCal pH Buffer 11.00 (Ref 238278)
A prerequisite for an accurate pH calibration is the availability of suitable buffer solutions. Hamilton 11.00 DuraCal liquid buffer solution (Ref 238278) has an accuracy of +/- 0.05 pH units and a stability of 24 months. The pH value of the buffer is affected by changing temperature. The effects of temperature from 5 to 50°C is shown below in 1 degree increments.
pH vs. Temperature: Reference table for 11.00 DuraCal buffer
Temp [°C] | pH Value | Temp [°C] | pH Value |
---|---|---|---|
5 | 11.33 | 30 | 10.93 |
6 | 11.32 | 31 | 10.91 |
7 | 11.30 | 32 | 10.90 |
8 | 11.28 | 33 | 10.89 |
9 | 11.26 | 34 | 10.87 |
10 | 11.25 | 35 | 10.86 |
11 | 11.23 | 36 | 10.85 |
12 | 11.21 | 37 | 10.83 |
13 | 11.19 | 38 | 10.82 |
14 | 11.17 | 39 | 10.81 |
15 | 11.16 | 40 | 10.80 |
16 | 11.14 | 41 | 10.78 |
17 | 11.12 | 42 | 10.77 |
18 | 11.11 | 43 | 10.76 |
19 | 11.09 | 44 | 10.74 |
20 | 11.07 | 45 | 10.73 |
21 | 11.06 | 46 | 10.72 |
22 | 11.03 | 47 | 10.71 |
23 | 11.03 | 48 | 10.70 |
24 | 11.02 | 49 | 10.68 |
25 | 11.00 | 50 | 10.67 |
26 | 10.99 | ||
27 | 10.97 | ||
28 | 10.96 | ||
29 | 10.94 | ||
30 | 10.93 |