Register for a Hamilton Webinar!

A prerequisite for an accurate pH calibration is the availability of suitable buffer solutions. Hamilton 11.00 DuraCal liquid buffer solution (Ref 238278) has an accuracy of +/- 0.05 pH units and a stability of 24 months. The pH value of the buffer is affected by changing temperature. The effects of temperature from 5 to 50°C is shown below in 1 degree increments.

pH vs. Temperature: Reference table for 11.00 DuraCal buffer

Temp [°C]pH ValueTemp [°C]pH Value
5 11.33 30 10.93
6 11.32 31 10.91
7 11.30 32 10.90
8 11.28 33 10.89
9 11.26 34 10.87
10 11.25 35 10.86
11 11.23 36 10.85
12 11.21 37 10.83
13 11.19 38 10.82
14 11.17 39 10.81
15 11.16 40 10.80
16 11.14 41 10.78
17 11.12 42 10.77
18 11.11 43 10.76
19 11.09 44 10.74
20 11.07 45 10.73
21 11.06 46 10.72
22 11.03 47 10.71
23 11.03 48 10.70
24 11.02 49 10.68
25 11.00 50 10.67
26 10.99
27 10.97
28 10.96
29 10.94
30 10.93