pH and ORP Cleaning / Storage Solutions
Downloads & FAQ
Hier finden Sie eine Reihe von themenspezifischen Artikeln, eine Auflistung häufig gestellter Fragen und zahlreiche Dokumente zum Herunterladen.
-
Catalog - Process Analytics
Download the Process Analytics sensor catalog to learn about the complete offering of Hamilton process sensors and accessories. -
SDS - Cleaning Solution B (237438)
-
SDS - Cleaning Solution A (237436)
-
What solution should I store my pH or ORP probe in?
Hamilton recommends storing pH and ORP probes wet in either Hamilton Storage Solution (Ref 238931) or in 3M KCl solution. Click to see the pH/ ORP Sensor Storage and Regeneration Knowledgebase for more information. -
How do I clean an ORP sensor?
The platinum band around the tip of the ORP sensor can oxidize over time. Light scrubbing with a mild abrasive such as toothpaste will remove the oxidization. Rinse with potable water and store in storage solution prior to usage -
How does the DuraCal buffer bottle work?
The Calpak bottle is a unique, patented design which includes a small chamber at the top of the bottle that fills with buffer solution when the bottle is squeezed. Click to see a simple pictorial of how to use this bottle.
-
Glossary of Sensor & Biopharma Acronyms & Abbreviations
Acronyms and abbreviations are commonly used with process analytics measurement found in the biopharma industry. This glossary is an attempt to de-code some of the more frequently used acronyms.
Wir sind hier um zu helfen!
Die Mitarbeiter des Vertriebs, des Kundendienstes und des technischen Supports von Hamilton stehen jederzeit für Fragen und zusätzliche Informationen zu allen Produkten und Dienstleistungen zur Verfügung