OxyGold B Traditional Sauerstoffsensoren
Die bewährten OxyGold B-Sensoren für gelösten Sauerstoff sind gut für Anwendungen geeignet, die Kohlendioxid oder andere saure Gase enthalten. Der Sensor bietet einen Standard-VP-Anschluss mit einem traditionellen Nanoampere (nA)-Ausgang zum Anschluss an die meisten handelsüblichen Controller und Transmitter.
Downloads & FAQ
-
Catalog - Process Analytics
Download the Process Analytics sensor catalog to learn about the complete offering of Hamilton process sensors and accessories. -
Brochure - Brewery
Learn about our control and analytical solutions that are essential to the brewing process. -
Manual - OxyGold B Sensors
Manual for OxyGold B Sensors
-
Does temperature impact my oxygen measurement?
Yes, in a closed system, gas phase temperature is directly proportional to pressure and therefore measured oxygen content. In solution, temperature is inversely proportional to dissolved oxygen content. -
When should I calibrate my DO sensor?
The frequency of calibration will depend upon the duration of process runs and the accuracy required for process control. In bioreactor applications calibration can be performed before sterilization; however many customers choose to calibrate after sterilization as well. -
When should I replace the membrane and electrolyte in my polarographic DO sensor?
The frequency of membrane and electrolyte replacement is process specific and will vary. Routine calibration before each run can identify a high nA current at the nitrogen zero point or low nA current in air, indicating a need to replace the membrane and electrolyte. Click to learn more about membrane replacement.