The Conducell 4USF Arc is a 4-pole conductivity sensor equipped with advanced Arc technology and a VP8 connector. This sensor is designed for hygienic applications, offering features like onboard calibration storage and direct connection to control systems, simplifying integration and maintenance.
Product Specifications
| Wetted Parts |
Wetted Parts
Stainless Steel 1.4435
PEEK (Polyetheretherketone) EPDM (Ethylene propylene elastomer) See compliance details in Material Specification document |
|---|---|
| Electrical Connector |
Electrical Connector
VP8
|
| Sensor Output |
Sensor Output
Arc: Modbus, 4 to 20 mA
|
| Measuring Range |
Measuring Range
1 µS/cm to 300 mS/cm
|
| Accuracy at 25 °C |
Accuracy at 25 °C
± 3 % at 1 µS/cm to 100 mS/cm, ± 5 % at 100 to 300 mS/cm
|
| Cell Constant |
Cell Constant
0.36 /cm
|
| Temperature Sensor |
Temperature Sensor
NTC 22 kOhm
|
| Configurable Values |
Configurable Values
Conductivity: µS/cm, mS/cm; Temperature: °C, °F, K
|
| Surface Quality of Steel |
Surface Quality of Steel
Ra < 0.4 µm (N5, electropolished)
|
| Analog Interface 1 |
Analog Interface 1
4 to 20 mA for conductivity, programmable
|
| Analog Interface 2 |
Analog Interface 2
4 to 20 mA for Temp., programmable
|
| Analog Interface 1 and 2 |
Analog Interface 1 and 2
Galvanically not isolated; pulse width modulation 3.5 kHz
|
| Digital Interface |
Digital Interface
RS485 Modbus (max. 32 addresses)
|
| Baud Rate |
Baud Rate
4800, 9600, 19200, 38400, 57600, 115200 Bd
|
| Operating Voltage |
Operating Voltage
7 to 30 VDC max. 150 mW
|
| Serial Number |
Serial Number
Yes
|
| Certificate |
Certificate
Yes, with calibrated cell constant
|
| ATEX Approval |
ATEX Approval
No
|
| IECEx Approval |
IECEx Approval
No
|
| Diameter |
Diameter
12 mm
|
| Autoclavable |
Autoclavable
Yes, max. Temperature 130 °C
|
| Steam Sterilizable |
Steam Sterilizable
Yes, max. Temperature 140 °C
|
| CIP |
CIP
Yes
|
| Operating Temperature Range |
Operating Temperature Range
Analog interface: 0 to 110 °C; Digital interface: 0 to 140 °C
|
| Pressure Range bar g |
Pressure Range bar g
0 to 20 bar (130 °C)
|
| Electrode Material |
Electrode Material
Stainless steel 1.4435
|
Spec. Version B