ZEUS® X1 air displacement technology employs Hamilton's innovative air displacement pipetting technology and CO-RE II® tips.
Product Specifications
| Accuracy |
Accuracy
<1% at 1000 µL, <5% at 1 µL
|
|---|---|
| Precision |
Precision
<0.75% at 1000 µL, <4% at 1 µL
|
| Weight |
Weight
1075 g
|
| Power Requirements |
Power Requirements
36.9-50 VDC, 2 A
|
| Communication Interface |
Communication Interface
CAN (Hamilton protocol), FCC connector
|
| Baud Rate |
Baud Rate
125,500 kBaud (others on request)
|
| Operating Temp High |
Operating Temp High
32 °C (90 °F)
|
| Operating Temp Low |
Operating Temp Low
18 °C (64 °F)
|
| Operating Humidity |
Operating Humidity
15% to 85% (no condensation, indoors)
|
| Storage Temp High |
Storage Temp High
70 °C (158 °F)
|
| Storage Temp Low |
Storage Temp Low
-25 °C (-13 °F)
|
| Storage Humidity |
Storage Humidity
15% to 85% (no condensation, indoors)
|
| z-Drive Excursion |
z-Drive Excursion
200 mm
|
| z-Drive Resolution |
z-Drive Resolution
0.1 mm
|
| z-Drive Maximum Speed |
z-Drive Maximum Speed
350 mm/sec
|
| Volume Range |
Volume Range
1 - 1,000 µL
|
| Pipetting Resolution |
Pipetting Resolution
0.1 µL
|
| Maximum Pipetting Speed |
Maximum Pipetting Speed
1,428 µL/s (peak speed)
|
| Pipetting Mode |
Pipetting Mode
Single and aliquot dispensing
|
| Tip Detection |
Tip Detection
Yes
|
| Liquid Level Detection |
Liquid Level Detection
Pressure-based (pLLD) and capacitive detection (cLLD)
|
| Pipetting Security |
Pipetting Security
Qualitative Pipette Monitoring (QPM) Anti-Droplet Control (ADC)
|
| Operating Noise |
Operating Noise
< 60 dBA, Indoor use only
|
| Device Addressing |
Device Addressing
up to 31 modules on a single CAN bus
|
| Dimensions (D x H x W) |
Dimensions (D x H x W)
161 mm x 335 mm x 16 mm
|
| Compatible CO-RE®II Tips |
Compatible CO-RE®II Tips
10 µL, 50 µL, 300 µL, 1 000 µL
|
| RoHS Compliant |
RoHS Compliant
Yes
|