The Incyte sensor measures viable cell permittivity, correlating to VCD, for real-time bioreactor control. It eliminates the need for manual offline sampling.
Product Specifications
| Process Connection |
Process Connection
PG13,5
|
|---|---|
| Wetted Parts |
Wetted Parts
Platinum
Stainless Steel 1.4435 PEEK (Polyetheretherketone) EPDM (Ethylene propylene elastomer) See compliance details in Material Specification document |
| Electrical Connector |
Electrical Connector
VP11
|
| Measuring Range |
Measuring Range
0 to 700 pF/cm, equivalent to 5 x 10^5 to 8 x 10^9 cells/mL (mammalian) or 5 to 200 g/L dry weight (bacteria and yeast)
|
| Surface Quality of Steel |
Surface Quality of Steel
Ra < 0.4 µm (N5)
|
| Certificate |
Certificate
Yes
|
| Diameter |
Diameter
12 mm
|
| Autoclavable |
Autoclavable
Yes, max. Temperature 130 °C
|
| Steam Sterilizable |
Steam Sterilizable
Yes, max. Temperature 135 °C
|
| CIP |
CIP
Yes
|
| Operating Temperature Range |
Operating Temperature Range
0 to 60 °C
|
| Conductivity Range |
Conductivity Range
0.5 to 10 mS/cm
|
| Pressure Range bar g |
Pressure Range bar g
0 to 12 bar
|
| a-length |
a-length
220 mm
|
Spec. Version B