Conducell 4UTF VP 60

Specification Sheet (Part/REF # 243590-4261)

Conductivity Sensor Conducell 4USF

The Traditional Conducell 4UTF VP is a 4-pole conductivity sensor with Titanium electrodes and a VP6 connector, designed for hygienic applications and compatible with CIP/SIP/autoclave procedures. It provides precise conductivity measurements across a broad range and requires a Hamilton H100 transmitter.

Product Specifications

Measurement Principle
Measurement Principle
4 Pole Contacting
Electrical Connector
Electrical Connector
VP6
Measuring Range
Measuring Range
1 µS/cm to 500 mS/cm
Cell Constant
Cell Constant
0.36 /cm
Temperature Sensor
Temperature Sensor
Pt1000
Surface Quality of Steel
Surface Quality of Steel
Ra < 0.4 µm (N5, electropolished)
Wetted Parts
Wetted Parts
PEEK (Polyetheretherketone)
Titanium
EPDM (Ethylene propylene elastomer)
See compliance details in Material Specification document
Process Connection
Process Connection
PG13,5
Serial Number
Serial Number
Yes
Certificate
Certificate
Yes, with nominal and actual cell constant
ATEX Approval
ATEX Approval
II 1/2 G Ex ia IIC T4/T5/T6 Ga/Gb; II 1/2 D ia IIIC T x °C Da/Db
IECEx Approval
IECEx Approval
II 1/2 G Ex ia IIC T4/T5/T6 Ga/Gb; II 1/2 D ia IIIC T x °C Da/Db
Diameter
Diameter
12 mm
Autoclavable
Autoclavable
Yes
Steam Sterilizable
Steam Sterilizable
Yes
CIP
CIP
Yes
Operating Temperature Range
Operating Temperature Range
-20 to 150 °C
Pressure Range bar g
Pressure Range bar g
0 to 20 bar (135 °C), 0 to 10 bar (150 °C)
Electrode Material
Electrode Material
Titanium
a-length
a-length
60 mm
Spec. Version A