FlexiFlow SL 10

Specification Sheet (Part/REF # 237340)

Flexiflow sensor housing

The FlexiFlow flow-through cell enables reliable pH and oxygen measurements, particularly in ion-weak media like coolant piping. Its compact design allows easy attachment, and it is self-draining.

Product Specifications

Process Connection
Process Connection
Swagelok 10 mm
Wetted Parts
Wetted Parts
Stainless Steel 1.4435
EPDM (Ethylene propylene elastomer)
(FDA 21 CFR 177.2600; EU 1935/2004; USP <88> Class VI - 121 °C)
See compliance details in Material Specification document
For Sensors with
For Sensors with
120 mm, 12 mm, PG13,5
Max. Diameter
Max. Diameter
65 mm
Max. Total Length
Max. Total Length
175 mm
Min. Total Length
Min. Total Length
175 mm
Internal Volume
Internal Volume
Approx. 5.5 mL
Mounting
Mounting
Bottom Screw Mount (4 pcs)
Weight
Weight
1200 g
Serial Number
Serial Number
Yes
Certificate
Certificate
Yes, 3.1 certificate with heat number
ATEX Approval
ATEX Approval
II 1/2 G Ex ia IIC T4/T5/T6 Ga/Gb
II 1/2 D Ex ia IIIC T x °C Da/Db
IECEx Approval
IECEx Approval
II 1/2 G Ex ia IIC T4/T5/T6 Ga/Gb
II 1/2 D Ex ia IIIC T x °C Da/Db
EAC Ex Approval
EAC Ex Approval
II 1/2 G Ex ia IIC T4/T5/T6 Ga/Gb
II 1/2 D Ex ia IIIC T x °C Da/Db
UKEx Approval
UKEx Approval
II 1/2 G Ex ia IIC T4/T5/T6 Ga/Gb
II 1/2 D Ex ia IIIC T x °C Da/Db
Diameter
Diameter
65 mm
Autoclavable
Autoclavable
No
Steam Sterilizable
Steam Sterilizable
Yes
CIP
CIP
No
Operating Temperature Range
Operating Temperature Range
-10 to 130 °C
Max. Pressure bar g
Max. Pressure bar g
16 bar
Product Family
Product Family
FlowCell
Specifications are subject to change without notice
https://www.hamiltoncompany.com/sensors/sensor-housings/flowcell/flexiflow
Spec. Version C