GlucoSense Membrane

Specification Sheet (Part/REF # 10190449-5)

Glucosense membrane

The GlucoSense Membrane is a key component for selective, in-situ glucose measurement. It acts as a molecular filter, allowing glucose to diffuse while excluding larger particles, such as cells. Essential for consistent, accurate readings in complex cell culture media.

Product Specifications

Wetted Parts
Wetted Parts
PEEK (Polyetheretherketone)
PTFE (Polytetrafluoroethylene)
Silicone
See compliance details in 2.1 Material Specification document REF 111011994
Biocompatibility
Biocompatibility
Yes, study report available upon request.
Shelf Life
Shelf Life
1 year from manufacturing date
Valid as long as the defined storage conditions are maintained.
Autoclavable
Autoclavable
Yes (130°C, 30-minute sterilization duration). The membrane can be autoclaved only once.
Storage Conditions
Storage Conditions
The GlucoSense Membranes must be kept refrigerated between 2°C to 8°C. Refrigeration is not required during shipment.
Operating Temperature Range
Operating Temperature Range
-10 °C to 130 °C
Protection Rating
Protection Rating
IP68
Required Flow
Required Flow
None
Specifications are subject to change without notice
https://www.hamiltoncompany.com/sensors/glucose-sensors/glucosense-membrane
Spec. Version A