FlowCell COND 4UPtF Arc

Specification Sheet (Part/REF # 10168201-xyz)

The FlowCell COND 4UPtF Arc is a high-precision conductivity sensor for inline process monitoring. Featuring a four-electrode platinum design, it provides accurate and stable measurements across a wide conductivity range. Hamilton’s Arc intelligent sensor technology enables direct digital communication to eliminate external transmitters and simplify system integration.

Product Specifications

Wetted Parts
Wetted Parts
Stainless Steel 1.4435<br>PEEK (Polyetheretherketone)<br>EPDM (Ethylene propylene elastomer)
O-ring Material
O-ring Material
EPDM (FDA approved, USP Class VI)
Pressure Range (bar g)
Pressure Range (bar g)
0 to 16 bar (140 °C)
Surface Quality of Steel
Surface Quality of Steel
Ra < 0.4 µm (N5, electropolished)
ATEX approval
ATEX approval
No
IECEx approval
IECEx approval
No
Process Connection
Process Connection
1/4" TC25 (ID 4.5 mm)<br>3/8" TC 25 (7.74 mm)<br>1/2" TC 25 (9.4 mm)<br>3/4" TC 25 (15.75 mm)<br>1" TC 50 (22.1 mm), not possible for internal diameter of 15 mm
Operating Temperature Range
Operating Temperature Range
Analog interface: 0 to 110 °C; Digital interface: 0 to 140 °C
Internal Volume
Internal Volume
15mm Internal Diameter: Approx. 20 mL<br>20mm Internal Diameter: Approx. 27mL<br>20mm Internal Diameter with pH: Approx. 37mL
Mounting
Mounting
2 Hex screws: M4*45 mm
Serial Number
Serial Number
Yes
Certificate
Certificate
Yes, with calibrated cell constant
CIP
CIP
Yes
Electrical Connector
Electrical Connector
VP8
Sensor Output
Sensor Output
Arc: Modbus, 4 to 20 mA
Measuring Range
Measuring Range
1 µS/cm to 300 mS/cm
Accuracy at 25 °C
Accuracy at 25 °C
± 3% or better from 1 μS/cm – 100 mS/cm<br>± 5% or better from 100 mS/cm - 300 mS/cm
Cell Constant
Cell Constant
0.4 / cm (FlowCell COND 4UPtF Arc)<br>0.415 / cm (FlowCell COND 4UPtF Arc + pH)
Temperature Sensor
Temperature Sensor
NTC 22 kOhm
Configurable Values
Configurable Values
Conductivity: µS/cm, mS/cm; Temperature: °C, °F, K
Analog Interface 1
Analog Interface 1
4 to 20 mA for conductivity, programmable
Analog Interface 2
Analog Interface 2
4 to 20 mA for Temp., programmable
Analog Interface 1 and 2
Analog Interface 1 and 2
Galvanically not isolated; pulse width modulation 3.5 kHz
Digital Interface
Digital Interface
RS485 Modbus (max. 32 addresses)
Baud Rate
Baud Rate
4800, 9600, 19200, 38400, 57600,115200 Bd
Operating Voltage
Operating Voltage
7 to 30 VDC max. 150 mW
Temperature Response
Temperature Response
≤ 30 sec (Note: Temperature response is highly dependent on the process conditions including flow rate, environment temperature)
Internal Diameter of FlowCell
Internal Diameter of FlowCell
10168201-241: 15 mm<br>10168201-111: 20 mm
SIP
SIP
Yes, max. Temperature 140 °C
Electrodes
Electrodes
Platinum
Specifications are subject to change without notice