Optical DO Cap S2 Kit

Specification Sheet (Part/REF # 10077858)

Optical DO Cap S2 Kit Single Use Cap

Hamilton's ODO Cap S2 Kit is an optical dissolved oxygen sensor cap for biopharma processes. The ODO Cap S2 Kit is designed to be mounted in single-use containers like bench-top vessels. It connects to the VisiFerm SU for reliable measurement in single-use applications.

Product Specifications

Measurement Principle
Measurement Principle
Fluorescence Quenching Luminophore (optical dissolved oxygen)
Measuring Range
Measuring Range
4 ppb to 25 ppm (DO)
0 to 62.85 %-vol
0 to 300 %-sat
Accuracy at 25 °C
Accuracy at 25 °C
± 1.0 %-sat (at 20 %-sat)
± 1.0 %-sat (at 100 %-sat)
± 2.5 %-sat (at 200 %-sat)
All three accuracy values are valid with calibration values provided on sensor label and after product calibration after irradiation and dry storage up to 37 months
Drift at 25 °C
Drift at 25 °C
< 1 % per week (of measurement value)
Wet-in Time
Wet-in Time
< 30 min (after dry storage)
Response Time at 25 °C
Response Time at 25 °C
t98% < 30 s at 25 °C (from 0%-sat to 100%-sat)
Shelf Life
Shelf Life
24 months before irradiation
37 months after irradiation
Wetted Parts
Wetted Parts
PA (Cycloaliphatic polyamide)
EPDM (Ethylene propylene elastomer)
VMQ (Silicone elastomer)
See compliance details in Material Specification document
Burst Pressure
Burst Pressure
Up to 10 barg
Operating Temperature Range
Operating Temperature Range
4 to 50 °C
Storage Temperature Range
Storage Temperature Range
4 to 40 °C in non-condensing environment (<85% r. H.)
Transport Temperature Range
Transport Temperature Range
-30 to 60 °C (according to ASTM D4169-13)
Sterilization
Sterilization
Gamma irradiation compatible (up to 50 kGy)
X-Ray irradiation compatible (up to 50 kGy)
Not validated for autoclavation, Sterilization In Place (SIP), nor Cleaning In Place (CIP)
Chemical Resistance
Chemical Resistance
See Chemical Resistance Guide
Spec. Version E