Arc Module Cond-P SU

Specification Sheet (Part/REF # 10071707)

Single Use Conductivity

The Arc Module Cond-P (in combination with the Conducell-P SU) enables precise conductivity measurement in single-use bags.

Product Specifications

Electrical Connector
Electrical Connector
VP8
Parameter
Parameter
Conductivity (4-Pole)
Sensor Output
Sensor Output
Arc: Modbus
Measuring Range
Measuring Range
100 µS/cm to 300 mS/cm
Accuracy at 25 °C
Accuracy at 25 °C
± 5 % at 100 µS/cm to 100 mS/cm
± 8 % at 100 mS/cm to 200 mS/cm
± 10 % at 200 mS/cm to 300 mS/cm
Temperature Sensor
Temperature Sensor
Yes (on sensor element: Conducell-P SU)
Configurable Values
Configurable Values
Conductivity, Temperature
Digital Interface
Digital Interface
RS485 Modbus (max. 32 addresses)
Baud Rate
Baud Rate
4800, 9600, 19200, 38400, 57600, 115200 Bd
Operating Voltage
Operating Voltage
7 to 30 VDC max. 150 mW
Serial Number
Serial Number
Yes
Certificate
Certificate
Yes
Autoclavable
Autoclavable
No
Steam Sterilizable
Steam Sterilizable
No
CIP
CIP
No
Gamma Irradiatable
Gamma Irradiatable
No
X-Ray irradiatable
X-Ray irradiatable
No
Operating Temperature Range
Operating Temperature Range
0 to 60 °C
Protection Rating
Protection Rating
IP64 (when connected to the sensor element: Conducell-P SU)
Spec. Version B