VisiTrace DO Ex (Discontinued)

Specification Sheet

Visitrace do ex Dissolved Oxygen Sensor

The discontinued VisiTrace DO Ex sensor family was designed for industrial environments with standard and Ex versions for use in explosive environments.

Product Specifications

Process Connection
Process Connection
PG13,5
Wetted Parts
Wetted Parts
Stainless Steel 1.4435
EPDM (Ethylene propylene elastomer)
Silicone - FDA 21 CFR 177.2600
VMQ (Silicone elastomer)
See compliance details in Material Specification document
Electrical Connector
Electrical Connector
M12
Measuring Range
Measuring Range
0 to 2 ppm (0 to 50 mbar)
Electrolyte
Electrolyte
None
Accuracy at 25 °C
Accuracy at 25 °C
± 0.5 ppb or 2 % whichever is greater
Drift at Room Temperature
Drift at Room Temperature
< 1 % per week @ 100 ppb
Sensor Cap
Sensor Cap
ODO Cap L0
Oxygen Consumption
Oxygen Consumption
None
Temperature Sensor
Temperature Sensor
NTC 22 kOhm
Configurable Values
Configurable Values
Dissolved Oxygen (DO): mbar; %-sat; %-vol; µg/l; mg/l; ppb/ppm (gas); ppb/ppm; Temperature: °C, °F, K
Surface Quality of Steel
Surface Quality of Steel
Ra < 0.4 µm (N5)
4 to 20 mA Current Range
4 to 20 mA Current Range
3.5 to 22 mA
Analog Interface 1
Analog Interface 1
Two wire sink 4 to 20 mA for DO, programmable
4 to 20 mA Accuracy
4 to 20 mA Accuracy
< 0.3 % + 0.05 mA
Transmitting Power
Transmitting Power
Max. 4 dBm
Digital Interface
Digital Interface
Bluetooth 4.0
Operating Voltage
Operating Voltage
18 to 30 VDC
Serial Number
Serial Number
Yes
Response Time t90%
Response Time t90%
< 20 s in gas; < 90 s in water
Certificate
Certificate
Yes, with measured current in oxygen
ATEX Approval
ATEX Approval
ll 1G Ex ia llC T6/T5/T4/T3 Ga; ll 1D Ex ia lllC T135°C Da
IECEx Approval
IECEx Approval
ll 1G Ex ia llC T6/T5/T4/T3 Ga; ll 1D Ex ia lllC T135°C Da
Diameter
Diameter
12 mm
Autoclavable
Autoclavable
No
Steam Sterilizable
Steam Sterilizable
Yes, max. Temperature 140 °C
CIP
CIP
Yes
Operating Temperature Range
Operating Temperature Range
-10 to 140 °C; the sensor provides no DO reading above 85 °C
Pressure Range bar g
Pressure Range bar g
-1 to 12 bar
Protection Rating
Protection Rating
IP68
Required Flow
Required Flow
None
Specifications are subject to change without notice
https://www.hamiltoncompany.com/sensors/visitrace-do-ex
Spec. Version B