The Hamilton GlucoSense Optical Sensor provides real-time, in-situ glucose monitoring essential for bioprocessing R&D. Its easy-to-operate design ensures accurate measurements from day one, optimizing mammalian cell culture productivity. Overcome your biopharma challenges with reliable glucose control.

Product Specifications

Wetted Parts
Wetted Parts
Titanium 3.7035 coated with diamond like carbon
Diamond
EPDM (Ethylene propylene elastomer)
See compliance details in 2.1 Material Specification document REF 111011993
Electrical Connector
Electrical Connector
VP8
Sensor Output
Sensor Output
Arc: Modbus
Measuring Range
Measuring Range
0.5 to 25.0 g/L
pH Value
pH Value
The specified glucose measurement accuracy applies when the process pH remains constant within ±0.2 units
Electrolyte
Electrolyte
None
Accuracy at 25 °C
Accuracy at 25 °C
±0.5 g/l after product calibration at the measurement temperature. Valid for whole measurement range.
Sensor Insertion Depth
Sensor Insertion Depth
Minimum 20 mm above the sensing element
Temperature Sensor
Temperature Sensor
NTC 47 kOhm
Configurable Values
Configurable Values
Glucose concentration: g/l, mg/dl, mg/l, mmol/l; Temperature: K, °C, °F
O-ring Material
O-ring Material
See compliance details in Material Specification document
Process Connection
Process Connection
PG13,5
Digital Interface
Digital Interface
RS485 Modbus (max. 32 addresses)
Operating Voltage
Operating Voltage
10 to 27 VDC
Certificate
Certificate
Yes, factory calibration with certified solutions
Diameter
Diameter
12 mm
Autoclavable
Autoclavable
Yes (130 °C, 30-minute sterilization duration).
Operating Temperature Range
Operating Temperature Range
-10 to 130 °C; the sensor provides no Glucose reading above 60 °C
Pressure Range bar g
Pressure Range bar g
0 to 5 bar(g)
Protection Rating
Protection Rating
IP68
Required Flow
Required Flow
None
Temperature Measuring Range
Temperature Measuring Range
The declared glucose measurement accuracy is valid within the temperature range from 25°C to 38°C.
Software
Software
ArcAir 3.11 or successive releases
Surface Quality of Titanium
Surface Quality of Titanium
Titanium coated with diamond like carbon with Ra < 0.4 μm (N5)
Specifications are subject to change without notice
https://www.hamiltoncompany.com/sensors/glucose-sensors/glucosense
Spec. Version A