The Hamilton GlucoSense Optical Sensor provides real-time, in-situ glucose monitoring essential for bioprocessing R&D. Its easy-to-operate design ensures accurate measurements from day one, optimizing mammalian cell culture productivity. Overcome your biopharma challenges with reliable glucose control.
Product Specifications
| Wetted Parts |
Wetted Parts
Titanium 3.7035 coated with diamond like carbon
Diamond EPDM (Ethylene propylene elastomer) See compliance details in 2.1 Material Specification document REF 111011993 |
|---|---|
| Electrical Connector |
Electrical Connector
VP8
|
| Sensor Output |
Sensor Output
Arc: Modbus
|
| Measuring Range |
Measuring Range
0.5 to 25.0 g/L
|
| pH Value |
pH Value
The specified glucose measurement accuracy applies when the process pH remains constant within ±0.2 units
|
| Electrolyte |
Electrolyte
None
|
| Accuracy at 25 °C |
Accuracy at 25 °C
±0.5 g/l after product calibration at the measurement temperature. Valid for whole measurement range.
|
| Sensor Insertion Depth |
Sensor Insertion Depth
Minimum 20 mm above the sensing element
|
| Temperature Sensor |
Temperature Sensor
NTC 47 kOhm
|
| Configurable Values |
Configurable Values
Glucose concentration: g/l, mg/dl, mg/l, mmol/l; Temperature: K, °C, °F
|
| O-ring Material |
O-ring Material
See compliance details in Material Specification document
|
| Process Connection |
Process Connection
PG13,5
|
| Digital Interface |
Digital Interface
RS485 Modbus (max. 32 addresses)
|
| Operating Voltage |
Operating Voltage
10 to 27 VDC
|
| Certificate |
Certificate
Yes, factory calibration with certified solutions
|
| Diameter |
Diameter
12 mm
|
| Autoclavable |
Autoclavable
Yes (130 °C, 30-minute sterilization duration).
|
| Operating Temperature Range |
Operating Temperature Range
-10 to 130 °C; the sensor provides no Glucose reading above 60 °C
|
| Pressure Range bar g |
Pressure Range bar g
0 to 5 bar(g)
|
| Protection Rating |
Protection Rating
IP68
|
| Required Flow |
Required Flow
None
|
| Temperature Measuring Range |
Temperature Measuring Range
The declared glucose measurement accuracy is valid within the temperature range from 25°C to 38°C.
|
| Software |
Software
ArcAir 3.11 or successive releases
|
| Surface Quality of Titanium |
Surface Quality of Titanium
Titanium coated with diamond like carbon with Ra < 0.4 μm (N5)
|
Spec. Version A