10 µL, Microliter Autosampler Syringe, Removable Needle (RN), PN: 80338

Specification Sheet (Part/REF # 80338)

10 µL, Microliter Autosampler Syringe, Removable Needle (RN), PN: 80338

Hamilton's 700 series, Microliter #701 Removable Needle (RN) Syringe, Model701, is designed for manual dispensing and use in autosamplers. This syringe is commonly used with autosamplers such as Agilent GC 7670, 7671, 7672. Hamilton lab syringes ensure an accuracy of ± 1%, superior durability, and are manufactured with the highest quality chemically resistant materials.

Product Specifications

Volume
Volume
10 µL
Syringe Type
Syringe Type
Microliter
Syringe Series
Syringe Series
700
Termination
Termination
Removable Needle (RN)
Needle
Needle
Small RN
Gauge
Gauge
26s gauge
Point Style
Point Style
2: Curved, Beveled
Needle Length
Needle Length
2 inch (51 mm)
Plunger Style
Plunger Style
Agilent Plunger
Fluid Path
Fluid Path
SST, borosilicate glass
Autoclavable
Autoclavable
Yes
Gas Sterilizable
Gas Sterilizable
Yes
Pressure
Pressure
2000 psig (137.9 bar)
Maximum Operating Temperature
Maximum Operating Temperature
80°C / 176°F
Minimum Operating Temperature
Minimum Operating Temperature
10°C / 50°F
Barrel Inner Diameter
Barrel Inner Diameter
0.019 inches (0.485 mm)
Barrel Outer Diameter
Barrel Outer Diameter
0.260 inches (6.60 mm)
Needle Inner Diameter
Needle Inner Diameter
0.005 inches (0.13 mm)
Needle Outer Diameter
Needle Outer Diameter
0.0187 inches (0.47 mm)
Std Needle Dead Volume
Std Needle Dead Volume
0.64 µL
Sales Quantity
Sales Quantity
1 syringe per pack
Syringe Model Number
Syringe Model Number
701
Specifications are subject to change without notice