The EasyFerm Bio MS2 pH sensors feature Memosens input and Hamilton pre-pressurized technology. Ideal for food and beverage applications.
Product Specifications
| Process Connection |
Process Connection
PG13,5
|
|---|---|
| Wetted Parts |
Wetted Parts
GΙass
Ceramic VMQ (Silicone elastomer) |
| Electrical Connector |
Electrical Connector
Memosens
|
| Measuring Range |
Measuring Range
pH 0 to 14
|
| Electrolyte |
Electrolyte
Pressurized Foodlyte
|
| Sensitivity |
Sensitivity
57 to 59 mV / pH at 25 °C
|
| Zero Point |
Zero Point
0 ± 20 mV
|
| pH Glass Resistance |
pH Glass Resistance
Moderate
|
| Membrane Shape |
Membrane Shape
Spherical
|
| Diaphragm |
Diaphragm
HP-Coatramic
|
| Number of Diaphragms |
Number of Diaphragms
1
|
| Reference System |
Reference System
Hamilton Everef-F, Silver ion barrier included
|
| Liquid Earth |
Liquid Earth
No
|
| Temperature Sensor |
Temperature Sensor
NTC 30 kOhm
|
| O-ring Material |
O-ring Material
FDA 21 CFR 177.2600; EC 1935/2004; USP <87>; USP <88> Class VI (121 °C)
|
| Serial Number |
Serial Number
Yes
|
| Recommended Storage |
Recommended Storage
Hamilton Storage Solution or 3 M KCl
|
| Certificate |
Certificate
Yes, with measured values in Buffer pH 4.01 and pH 7.00
|
| ATEX Approval |
ATEX Approval
II 1 G Ex ia IIC T3/T4/T6 Ga
|
| IECEx Approval |
IECEx Approval
II 1 G Ex ia IIC T3/T4/T6 Ga
|
| Diameter |
Diameter
12 mm
|
| Autoclavable |
Autoclavable
Yes
|
| Steam Sterilizable |
Steam Sterilizable
Yes
|
| CIP |
CIP
Yes
|
| Operating Temperature Range |
Operating Temperature Range
0 to 140 °C
Explosive Atmospheres: See Operating Instructions |
| Sample min Conductivity |
Sample min Conductivity
100 µS/cm
|
| Pressure Range bar g |
Pressure Range bar g
0 to 6 bar
|
| pH Glass |
pH Glass
Hamilton type HB glass
|
| a-length |
a-length
120 mm
|
Spec. Version B