Conducell 4UPtF Arc 30

Specification Sheet (Part/REF # 243590-2151)

Conductivity Sensor - Conducell 4USF Arc

The Conducell 4UPtF Arc is a 4-pole conductivity sensor featuring Arc technology for enhanced data management and direct control system connectivity via a VP8 connector and PG13,5 process connection. Its Platinum electrodes ensure long-term stability and corrosion resistance, making it ideal for hygienic applications with CIP/SIP/autoclave compatibility.

Product Specifications

Process Connection
Process Connection
PG13,5
Wetted Parts
Wetted Parts
Platinum
PEEK (Polyetheretherketone)
EPDM (Ethylene propylene elastomer)
See compliance details in Material Specification document
Electrical Connector
Electrical Connector
VP8
Sensor Output
Sensor Output
Arc: Modbus, 4 to 20 mA
Measuring Range
Measuring Range
1 µS/cm to 300 mS/cm
Accuracy at 25 °C
Accuracy at 25 °C
± 3 % at 1 µS/cm to 100 mS/cm, ± 5 % at 100 to 300 mS/cm
Cell Constant
Cell Constant
0.36 /cm
Temperature Sensor
Temperature Sensor
NTC 22 kOhm
Configurable Values
Configurable Values
Conductivity: µS/cm, mS/cm; Temperature: °C, °F, K
Surface Quality of Steel
Surface Quality of Steel
Ra < 0.4 µm (N5, electropolished)
Analog Interface 1
Analog Interface 1
4 to 20 mA for conductivity, programmable
Analog Interface 2
Analog Interface 2
4 to 20 mA for Temp., programmable
Analog Interface 1 and 2
Analog Interface 1 and 2
Galvanically not isolated; pulse width modulation 3.5 kHz
Digital Interface
Digital Interface
RS485 Modbus (max. 32 addresses)
Baud Rate
Baud Rate
4800, 9600, 19200, 38400, 57600, 115200 Bd
Operating Voltage
Operating Voltage
7 to 30 VDC max. 150 mW
Serial Number
Serial Number
Yes
Certificate
Certificate
Yes, with calibrated cell constant
ATEX Approval
ATEX Approval
No
IECEx Approval
IECEx Approval
No
Diameter
Diameter
12 mm
Autoclavable
Autoclavable
Yes, max. Temperature 130 °C
Steam Sterilizable
Steam Sterilizable
Yes, max. Temperature 140 °C
CIP
CIP
Yes
Operating Temperature Range
Operating Temperature Range
Analog interface: 0 to 110 °C; Digital interface: 0 to 140 °C
Pressure Range bar g
Pressure Range bar g
0 to 20 bar (130 °C)
Electrode Material
Electrode Material
Platinum
a-length
a-length
30 mm
Spec. Version B