Polilyte Plus H Arc 225

Specification Sheet (Part/REF # 242428-1323)

pH sensor Polilyte Plus Arc

The Polilyte Plus Arc is a maintenance-free, intelligent pH sensor with a long lifetime, designed to withstand demanding industrial applications.

Product Specifications

Process Connection
Process Connection
PG13,5
Wetted Parts
Wetted Parts
GΙass
FKM
Electrical Connector
Electrical Connector
VP8
Measuring Range
Measuring Range
pH 0 to 14
Electrolyte
Electrolyte
Polisolve Plus
Sensitivity
Sensitivity
57 to 59 mV / pH at 25 °C
Zero Point
Zero Point
0 ± 20 mV
pH Glass Resistance
pH Glass Resistance
Moderate
Membrane Shape
Membrane Shape
Spherical
Diaphragm
Diaphragm
Single Pore
Number of Diaphragms
Number of Diaphragms
2
Reference System
Reference System
Hamilton Everef-L
Liquid Earth
Liquid Earth
No
Temperature Sensor
Temperature Sensor
NTC 22 kOhm
Configurable Values
Configurable Values
pH: pH, mV; Temperature: °C, K, °F
Analog Interface 1
Analog Interface 1
4 to 20 mA for pH, programmable
Analog Interface 2
Analog Interface 2
4 to 20 mA for Temp., programmable
Analog Interface 1 and 2
Analog Interface 1 and 2
Galvanically not isolated; pulse width modulation 3.5 kHz
Digital Interface
Digital Interface
RS485 Modbus (max. 32 addresses)
Baud Rate
Baud Rate
4800, 9600, 19200, 38400, 57600, 115200 Bd
Operating Voltage
Operating Voltage
7 to 30 VDC max. 150 mW
Serial Number
Serial Number
Yes
Recommended Storage
Recommended Storage
Hamilton Storage Solution or 3 M KCl
Certificate
Certificate
Yes, with measured values in Buffer pH 4.01 and pH 7.00
ATEX Approval
ATEX Approval
No
IECEx Approval
IECEx Approval
No
Diameter
Diameter
12 mm
Sample min Conductivity
Sample min Conductivity
2 µS/cm
Upside Down Installation
Upside Down Installation
Not Supported
pH Glass
pH Glass
Hamilton Type H Glass
a-length
a-length
225 mm
Autoclavable
Autoclavable
Yes, max. Temperature 130 °C
Steam Sterilizable
Steam Sterilizable
Yes, max. Temperature 130 °C
CIP
CIP
No
Operating Temperature Range
Operating Temperature Range
Analog interface: 0 to 100 °C; Digital interface: 0 to 130 °C
Pressure Range bar g
Pressure Range bar g
0 to 16 bar (100 °C), 0 to 10 bar (130 °C)
Spec. Version B