FlushTrode

Specification Sheet (Part/REF # 238060)

Hamilton Flushtrode pH electrode with blue cap and blue tip on transparent background

FlushTrode pH sensors provide low-maintenance, high-precision, and clog-resistant pH monitoring in tough applications. Their flush membrane, self-cleaning capability, and durability make these pH probes ideal for laboratories requiring hygienic, precise, and reliable pH measurement.

Product Specifications

Sensor Family
Sensor Family
FlushTrode
Wetted Parts
Wetted Parts
GΙass
FEP (Fluorinated Ethylene Propylene)
pH Glass
pH Glass
Hamilton Type HF Glass
Electrical Connector
Electrical Connector
S7
a-length
a-length
120 mm
Parameter
Parameter
pH
Measuring Range
Measuring Range
pH 0 to 14
Electrolyte
Electrolyte
3 M KCl
Sensitivity
Sensitivity
57 to 59 mV / pH at 25 °C
Zero Point
Zero Point
0 ± 20 mV
pH Glass Resistance
pH Glass Resistance
Moderate
Membrane Shape
Membrane Shape
Spherical
Diaphragm
Diaphragm
Sleeve
Number of Diaphragms
Number of Diaphragms
1
Reference System
Reference System
Hamilton Everef
Temperature Sensor
Temperature Sensor
No
Serial Number
Serial Number
Yes
Recommended Storage
Recommended Storage
Hamilton Storage Solution or 3 M KCl
Certificate
Certificate
Yes, with measured values in Buffer pH 4.01 and pH 7.00
Diameter
Diameter
12 mm
Operating Temperature Range
Operating Temperature Range
-10 to 80 °C
Pressure Range bar g
Pressure Range bar g
Atmospheric pressure
Spec. Version B