Hamilton’s Conducell-PO SU is a sensor element for biopharma processes. The Conducell-PO SU is designed to be pre-installed as an integral part of a single-use bag. It is optimized for welding to various bag films. It connects to the Arc Module Cond SU for reliable measurement in single-use applications.
Product Specifications
| Measurement Principle |
Measurement Principle
4 Pole Contacting
|
|---|---|
| Measuring Range |
Measuring Range
100 µS/cm to 300 mS/cm
|
| Accuracy at 25 °C |
Accuracy at 25 °C
± 5 % at 100 µS/cm to 100 mS/cm
± 8 % at 100 mS/cm to 200 mS/cm ± 10 % at 200 mS/cm to 300 mS/cm |
| Drift at 25 °C |
Drift at 25 °C
No observable drift
|
| Wet-in Time |
Wet-in Time
< 30 min (after dry storage)
|
| Response Time at 25 °C |
Response Time at 25 °C
t98% < 30 s (1.413 to 300 mS/cm)
|
| Shelf Life |
Shelf Life
24 months before irradiation
37 months after irradiation |
| Patch type |
Patch type
Overmold (See drawing)
|
| Sensor Output |
Sensor Output
Electrical resistivity
NTC 22 kOhm (temperature resistor) |
| Wetted Parts |
Wetted Parts
Platinum
Glass / Ceramic HDPE (High Density Polyethylene) TPE-O (LDPE) (Thermoplastic Elastomer Olefinic Base) See compliance details in Material Specification document |
| Burst Pressure |
Burst Pressure
Up to 10 barg
|
| Operating Temperature Range |
Operating Temperature Range
4 to 50 °C
|
| Storage Temperature Range |
Storage Temperature Range
4 to 40 °C in non-condensing environment (<85% r. H.)
|
| Transport Temperature Range |
Transport Temperature Range
-30 to 60 °C (according to ASTM D4169-13)
|
| Sterilization |
Sterilization
Gamma irradiation compatible (up to 50 kGy)
X-Ray irradiation compatible (up to 50 kGy) Not validated for autoclavation, Sterilization In Place (SIP), nor Cleaning In Place (CIP) |
| Chemical Resistance |
Chemical Resistance
See Chemical Resistance Guide
|
Spec. Version C