Incyte-PO SU

Specification Sheet (Part/REF # 10187043)

Incyte P SU - conductivity probe

Hamilton’s Incyte-PO SU is a sensor element for biopharma processes. The Incyte-PO SU is designed to be pre-installed as an integral part of a single-use stirred or wave bag. It is optimized for welding to various bag films. It connects to the Arc Module Incyte SU for reliable measurement in single-use applications.

Product Specifications

Measurement Principle
Measurement Principle
Permittivity measurement of viable cell volume
Measuring Range
Measuring Range
Permittivity: 0 to 700 pF/cm, equivalent to 5 x 10^5 to 8 x 10^9 cells/mL (mammalian)
Conductivity: 1 to 50 mS/cm
Precision at 25 °C
Precision at 25 °C
Permittivity (at 1 MHz): ± 1 pF or ± 1 %
Whichever value is greater over the entire measuring range
Accuracy at 25 °C
Accuracy at 25 °C
Conductivity (at 0 pF): ± 3 %
Over the entire measuring range
Drift at 25 °C
Drift at 25 °C
No observable drift
Wet-in Time
Wet-in Time
< 30 min (after dry storage)
Response Time at 25 °C
Response Time at 25 °C
t98% < 30 s (conductivity: 1.413 to 50 mS/cm)
Shelf Life
Shelf Life
24 months before irradiation
37 months after irradiation
Patch type
Patch type
Overmold (See drawing)
Sensor Output
Sensor Output
Electrical resistivity
Permittivity (Viable Cell Density)
NTC 22 kOhm (temperature resistor)
Wetted Parts
Wetted Parts
Platinum
Glass / Ceramic
HDPE (High Density Polyethylene)
TPE-O (LDPE) (Thermoplastic Elastomer Olefinic Base)
See compliance details in Material Specification document
Burst Pressure
Burst Pressure
Up to 10 barg
Operating Temperature Range
Operating Temperature Range
4 to 50 °C
Storage Temperature Range
Storage Temperature Range
4 to 40 °C in non-condensing environment (<85% r. H.)
Transport Temperature Range
Transport Temperature Range
-30 to 60 °C (according to ASTM D4169-13)
Sterilization
Sterilization
Gamma irradiation compatible (up to 50 kGy)
X-Ray irradiation compatible (up to 50 kGy)
Not validated for autoclavation, Sterilization In Place (SIP), nor Cleaning In Place (CIP)
Chemical Resistance
Chemical Resistance
See Chemical Resistance Guide
Spec. Version B