Hamilton’s Incyte-PO SU sensor elements are pre-installed as an integral part of a single-use container with a long shelf life. They provide accurate, stable measurements and come ready-to-use and pre-calibrated. They connect to the Arc Module Incyte SU for reliable measurement in single-use applications. Our SU VCD sensors combine the reliability and measurement stability of conventional sensors with the ease of use as an integral part of the bioreactor. Collect real-time viable cell density measurements for deeper insights and data-driven process optimization.
Product Specifications
| Wetted Parts |
Wetted Parts
Platinum
Ceramic / Glass HDPE (High Density Polyethylene) TPE-O (LDPE) (Thermoplastic Elastomer Olefinic Base) See compliance details in Material Specification document |
|---|---|
| Parameter |
Parameter
Viable Cell Density
|
| Measuring Range |
Measuring Range
0 to 700 pF/cm, equivalent to 5 x 10^5 to 8 x 10^9 cells/mL (mammalian)
|
| Calibration Data |
Calibration Data
Yes, provided on sensor label
|
| Temperature Sensor |
Temperature Sensor
Yes
|
| Biocompatibility |
Biocompatibility
Yes, see separate study report (upon request)
|
| Wet-in Time |
Wet-in Time
30 min after dry storage
|
| Shelf Life |
Shelf Life
12 months prior gamma irradiation
24 months after gamma irradiation |
| Serial Number |
Serial Number
No
|
| Certificate |
Certificate
Yes
|
| Autoclavable |
Autoclavable
No
|
| Steam Sterilizable |
Steam Sterilizable
No
|
| CIP |
CIP
No
|
| Gamma Irradiatable |
Gamma Irradiatable
Yes, up to 50 kGy
|
| X-Ray irradiatable |
X-Ray irradiatable
Yes, up to 50 kGy
|
| Operating Temperature Range |
Operating Temperature Range
4 to 50 °C
|
| Conductivity Range |
Conductivity Range
1 to 50 mS/cm
|
| Pressure Range bar g |
Pressure Range bar g
0 to 1 bar
|
| pH Range |
pH Range
3 to 10
|
| Patch type |
Patch type
Overmold
|
Spec. Version A