Specification Sheet (Part/REF # 10155129)

Ph Insert connector Single Use sensor Connection

Hamilton's pH Insert is a single-use armature for the assembly of the OneFerm pH 70 mm sensor into a 1" barb port for applications in single-use bags.

Product Specifications

Shelf Life
Shelf Life
18 months before irradiation
37 months after irradiation
Wetted Parts
Wetted Parts
PA (Cycloaliphatic polyamide)
See compliance details in Material Specification document
Burst Pressure
Burst Pressure
Up to 10 barg
Process Connection
Process Connection
PG13,5
Torque
Torque
1.5 Nm (of the OneFerm pH sensor)
Operating Temperature Range
Operating Temperature Range
4 to 50 °C
Storage Temperature Range
Storage Temperature Range
4 to 40 °C in non-condensing environment (<85% r. H.)
Transport Temperature Range
Transport Temperature Range
-30 to 60 °C (according to ASTM D4169-13)
Sterilization
Sterilization
Gamma irradiation compatible (up to 50 kGy)
X-Ray irradiation compatible (up to 50 kGy)
Not validated for autoclavation, Sterilization In Place (SIP), nor Cleaning In Place (CIP)
Chemical Resistance
Chemical Resistance
See Chemical Resistance Guide